Oscilloscope Utility Settings
6
Agilent 1000B Series Oscilloscopes User’s Guide
139
To set up masks
You can create masks by adding horizontal and vertical margins to a
signal. You can save and load masks from internal memory or an external
USB drive. And you can export and import masks from an exteral USB
drive.
To access the Mask menu:
1
Press
[Utility]
.
2
In the Utilities menu, press
Mask Test
.
3
In the Mask Test menu, press
MaskSetting
.
To adjust a mask’s horizontal failure margin
1
In the Mask menu (
[Utility]
>
Mask Test
>
MaskSetting
), press
X Mask
.
2
Turn the
entry knob to adjust the horizontal failure margin.
The margin can be set from 0.04 div to 4.00 div.
Figure 57
Mask Test Mask Setting
Содержание 1000 Series
Страница 1: ...s1 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 12: ...12 Agilent 1000B Series Oscilloscopes User s Guide Contents Safety Symbols 150 Index 151 ...
Страница 16: ...16 Agilent 1000B Series Oscilloscopes User s Guide Tables ...
Страница 36: ...1 Getting Started 36 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 53: ...Displaying Data 2 Agilent 1000B Series Oscilloscopes User s Guide 53 Figure 21 Waveform After Inversion ...
Страница 64: ...2 Displaying Data 64 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 98: ...3 Capturing Data 98 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 112: ...4 Making Measurements 112 Agilent 1000B Series Oscilloscopes User s Guide Figure 46 Cursors Key ...
Страница 116: ...4 Making Measurements 116 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 132: ...6 Oscilloscope Utility Settings 132 Agilent 1000B Series Oscilloscopes User s Guide Figure 53 Utility Key ...