DD+DIS033.07E
Repair and Service
Tools and Auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3.3.5
Testimage LinCheck
Purpose
Test image to check density reproduction.
Usage
(2) In the browser pages select menu
'Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage – 1 Test Image -
LinCheck.ZIP'.
(3) The test image as displayed in
Figure 14 will be printed.
536632ga.cdr
Figure 14
Evaluation
(4) The test image contains a grey
wedge. This grey wedge is
measured during the density
reproduction check.
Refer to chapter 3.6, Adjustments.
536632gb.cdr
Density measurement area
Figure 15
NOTE:
For check of density reproduction only measure in the density measurement area, as
there also the CDM is measuring.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.2 / 22
10-2007 Type
5367/100
Agfa Company Confidential