DD+DIS033.07E
Repair and Service
Tools and Auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3.3.4
Testimage Medical_Image
Purpose
Test image to check, whether print stripes or dots, which are slightly visible in one of
the technical images, are also visible in a medical test image.
Usage
(1) In the browser pages select menu
'Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage – 1 Test Image -
Medical_image.ZIP'.
(2) The test image as displayed in
Figure 13 will be printed.
Figure 13
Evaluation
Evaluate the test image at the lightbox as follows:
(1) This medical test image of a chest is the most critical medical test image
regarding visibility of print stripes or dots, due to its huge bright areas. In bright
areas the eye is most sensitive.
Check: Stripes in horizontal or vertical direction, which are slightly visible in a
technical image (e.g. STI test image), may not be visible in the medical test
image.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.2 / 21
10-2007 Type
5367/100
Agfa Company Confidential
Содержание DRYSTAR AXYS
Страница 7: ......
Страница 98: ......
Страница 106: ......
Страница 250: ......
Страница 290: ......
Страница 314: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 332: ......
Страница 334: ......
Страница 352: ......
Страница 372: ......
Страница 378: ......
Страница 409: ......
Страница 440: ......
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Страница 456: ......