Positioning
Programs
Generator
STITCHING
Cancel
Set
0
56
156
ROI Top
ROI Bottom
OID
Figure 9: Values for region of interest on DX-D 300
• Top: read the top of the region of interest on the vertical ruler of the
DX Full Leg Full Spine Stand. This is the start position.
• Bottom: read the bottom of the region of interest on the vertical ruler
of the DX Full Leg Full Spine Stand. This is the end position.
• Distance/OID: if the image is used for making length measurements in
NX or on true size printed images, read the distance between the
stitching grid of the DX Full Leg Full Spine Stand and the plane in
which measurements are to be made from the horizontal ruler on the
handle of the DX Full Leg Full Spine Stand. This distance is used for
calibrating distance measurements on the image in NX. If no
calibration is required, enter zero.
Note
:
The system imposes limits to the bottom and the top of
the region of interest, depending on the X-Ray system
installation.
3.
Set the X-Ray tube angle to 0 degrees.
4.
On the collimator, switch on the light localizer.
If required, collimate in transversal direction.
Instruction
:
For automatic stitching, the width of the
collimated area must be more than 15 cm.
Related Links
DR Full Leg Full Spine | Getting started |
33
0179E EN 20210708 1527