Symptom
Cause
Remedy
The IR Data Commu-
nication Unit is con-
nected to the wrong
USB port.
Reconnect the IR Data Com-
munication Unit to the USB
port where it was connected
during setup.
Battery thermal protection
The battery pack has a thermal protection that shuts down the battery power
in case of very high temperature.
Two situations can occur:
• Software protection: the battery pack will be usable as soon as the
temperature drops below a specified limit.
• Hardware protection: the battery pack must be exchanged.
DR 14s | Problem solving |