Warning
: The use of Agfa CR Phosphor Plate Cleaner and
Polynit wipes for CR MM3.xR is restricted to image plates
that have a batch number starting with the letter “C” or
higher. For CR MM3.xR image plates with a batch code
starting with the letter “B”, use only the dedicated
mammography PROSAT wipes purchased from Agfa or an
Agfa certified dealer to clean the plate.
Note
:
In a mixed environment with old and new CR MM3.xR plates
it is recommended to use only PROSAT wipes purchased from Agfa
or an Agfa certified dealer to clean the plates.
How to clean the image plate?
To clean the image plate:
1.
Insert a key or a pen into the cassette lock to open the shutter.
2.
Remove the key or the pen.
3.
Turn the cassette around, so that the black tube side is above.
4.
Hold back the shutter with both thumbs and let the image plate slide out
carefully onto the table.
5.
Put the image plate onto the black tube side of the empty cassette, as
shown in the figure below.
By putting the plate on the cassette so that the hooks are hanging over the
border of the cassette, a bending of the image plate is avoided.
Cleaning the image plate of CR HD5.0, CR MD4.0R, CR HM5.0 and CR MM3.0R | Cleaning the
Image Plate |
3
0042A EN 20170718 1541