Reading an image plate
CR 35-X D
IGITIZER
4
Remove the cassette from the cassette slot.
Reading an image plate in the dedicated
configuration
In the dedicated configuration, the CR 35-X can digitize the image plate while
you are entering the identification data via the ID Station.
To read an image plate:
1
Check that the CR 35-X is ready for operation:
•
The CR 35-X must display the operator main screen with ‘Ready’ status, e.g.:
•
The status indicator at the top of the CR 35-X must be green and be lit constantly.
2
Press the appropriate emergency button at the front of
the
CR 35-X.
The emergency button determines the speed class, i.e. the sensitivity, which will be
used to digitize the image plate.
WARNING:
When the CR 35-X returns the cassette, it is ready to be re-
used immediately. However, if you leave it for more than 3 days before
re-using it, you must re-erase it first. Refer to
on page
.
Note:
If the
CR 35-X
displays an error message, refer to
on page
Note: The CR 35-X is operational if the status field equals ‘READY’, even if
status messages of the destination are shown (e.g. ‘QS not ready’).
QS not ready
READY
Status field
Содержание CR 35-X
Страница 1: ...CR 35 X User Manual 4454C EN 2013 05 14 ...
Страница 4: ...4 4454C EN 2013 05 14 CR 35 X DIGITIZER ...
Страница 40: ...40 4454C EN 2013 05 14 Switching off the CR 35 X CR 35 X DIGITIZER ...
Страница 70: ...70 4454C EN 2013 05 14 Troubleshooting checklist CR 35 X DIGITIZER ...
Страница 71: ...Equipment information sheet Appendix A ...
Страница 77: ...CR cassettes Appendix B ...
Страница 88: ...88 4454C EN 2013 05 14 Technical specifications of the image plates CR 35 X DIGITIZER ...
Страница 89: ...Remarks for HF emission and immunity Appendix C ...
Страница 95: ...95 4454C EN 2013 05 14 Remarks for HF emission and immunity CR 35 X DIGITIZER ...
Страница 96: ...Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 4454C EN 2013 05 14 0413 ...