Starting DX-D 300
Note
:
Allow the DR Detector to warm up before the DX-D 300 is
used for clinical purposes. The warming-up time starts as soon as
the DR Detector has been powered on and the NX workstation is
running. To check if a warming-up time is required, refer to the DR
Detector User Manual.
1.
Switch on the X-Ray generator control in the operator room.
The X-Ray generator and the wired DR Detector are powered on.
2.
Switch on the U-arm unit using the button on the U-arm control box in the
examination room.
The U-arm unit and the touch screen console are powered on.
3.
Start NX.
The NX application and the software console are available on the NX
workstation.
For detailed information about starting up NX, refer to the NX User
Manual, document 4420.
4.
In a configuration with a wireless DR Detector, power on the DR Detector:
• attach a fully charged battery pack to the DR Detector.
• turn on the DR Detector.
• if needed, register the DR Detector to the NX workstation.
For detailed information about starting up the DR Detector, refer to the DR
Detector User Manual.