Equipment Classification
Per EN/IEC60601-1, Medical Electrical Equipment, General Requirements for
Safety 3rd Edition, this device is classified as following:
Table 52: Equipment classification
Class I equip-
ment
Equipment in which protection against electric shock does
not relay on basic insulation only, but includes a power sup-
ply cord with protective earth conductor. For earth reliability
always plug the main power cord into an earthed mains
power outlet.
Type B equip-
ment
A Type B piece of equipment is one that provides a particular
degree of protection against electric shock particularly re-
garding allowable leakage current and reliability of the pro-
tective earth protection.
Water ingress IPX0
This device does not have protection against ingress of wa-
ter.
Cleaning
See section on cleaning and disinfecting.
Disinfection
See section on cleaning and disinfecting.
Flammable
anesthetics
The equipment is not type AP or APG.
This device is not suitable for use in the presence of a flam-
mable anesthetic mixture with air, or in presence of a flam-
mable anesthetic mixture with oxygen or nitrous oxide.
Operation
Continuous operation with intermittent load.
Topics:
•
•
278
| DR 800 | Product Information
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...