Starting the DR Detector
CAUTION:
Do not use the battery pack as a power source for equipment
other than DX-D 45C or DX-D 45G detectors. Be sure to use only
the dedicated battery pack for the DX-D 45C or DX-D 45G
detector.
To start the DR Detector:
1.
If the DR Detector is connected to the System Control Unit via the DR
Detector cable, check if the power cable of the System Control Unit is
connected to the mains power.
The battery is not required to operate the DR Detector. Skip to step 4.
2.
Fully charge the battery.
Charge the battery on the day of examination or on the previous day.
Note
:
The battery slowly discharges even if not in use. The
battery pack may have expired if it discharges immediately
after being fully charged. You can purchase an optional battery
pack to replace an exhausted one.
3.
Attach the battery.
a) Open the cover of the battery slot
b) Slide the battery into the battery slot until it clicks.
The orientation of the battery is indicated at the bottom of the DR
Detector.
DX-D 45C, DX-D 45G | Getting started |
73
0292B EN 20190401 1627