Exposure blocked after switching to CR
Details
The exposure is set to CR using the DR Detector Switch. Expo-
sure is blocked.
Cause
The X-ray generator console is not automatically set to
free ex-
posure
. This situation applies only to Siemens generators.
Brief Solu-
tion
On the X-ray generator console, select
free exposure
. Perform
the CR exposure.