Establishing the reference values and verifying image quality
After installation of a new printer and before initial use you must establish
Quality Control aim values. These values will be used as the base line for
comparison when daily Quality Control is done. These values should be
determined again after major service, repair or software update.
The following Quality Control aim values must be determined:
• The daily operating density levels.
• Image geometry.
Once Quality Control aim values are established you must evaluate the Spatial
Resolution, the Artifact Levels and the Low Contrast Visibility to determine if
the image quality is acceptable.
The Quality Control aim values, the Spatial Resolution and Artifact Levels and
the Image Geometry values are all recorded on the Quality Control charts.
On these charts, the following test conditions are also recorded:
• The type and serial number of the printer.
• The type and emulsion number of the film used to determine the reference
values.
• The type of densitometer used.
• The time (day, month, year) that the values were established.
WARNING:
Before you can establish the daily operating levels, the printer
must be switched on for at least 15 minutes and it must be
calibrated as well.
Related Links
Establishing the daily operating reference density levels
on page 97
Establishing the image geometry reference values
on page 100
Verifying Acceptable Spatial Resolution, Artifact Levels and Low Contrast
on page 101
on page 124
on page 68
94
| Crystal Imager | Advanced Operation (key-operator mode)
3831B EN 20210601 1655
Содержание 5366/300
Страница 1: ...Crystal Imager 5366 300 User Manual 3831B EN 20210601 1655 ...
Страница 12: ...12 Crystal Imager Crystal Imager 3831B EN 20210601 1655 ...
Страница 57: ... The display Crystal Imager Introduction 57 3831B EN 20210601 1655 ...
Страница 89: ...WARNING Never reuse a jammed film Crystal Imager Basic Operation operator mode 89 3831B EN 20210601 1655 ...
Страница 114: ...5 Press the OK button on the touchscreen 114 Crystal Imager Preventive maintenance schedule 3831B EN 20210601 1655 ...
Страница 120: ... Cables transducers and accessories 120 Crystal Imager Remarks for HF emission and immunity 3831B EN 20210601 1655 ...
Страница 132: ...4 18 2832E EN 20151008 5 Remove plastic bag 132 Crystal Imager Plug Play Installation Manual 3831B EN 20210601 1655 ...
Страница 141: ...Crystal Imager Plug Play Installation Manual 141 3831B EN 20210601 1655 ...