accordance with
IEC 61000-4-8
the typical val-
ues as they are
in a
commercial and
clinical environ-
ment.
REMARK : U
r
is the alternating current in the network before the application
of the test level.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resist-
ance to Dis-
ruption
IEC 60601-1-2:2014
Test Level
Level of
Agree-
ment
Electromagnetic En-
vironment
Use portable and mo-
bile radio sets at a
safe distance from the
device (including the
leads) not closer than
the recommended
protective distance,
which is calculated
according to the equa-
tion suitable for the
RF emission frequen-
cy.
Recommended pro-
tective distance:
Conducted
high frequency
disturbance
variables in ac-
cordance with
IEC 61000-4-6
6 V/m in ISM fre-
quency bands be-
tween 150 kHz and
80 MHz
6 V/m in
ISM fre-
quency
bands be-
tween 150
kHz and 80
MHz
d
= 1.2
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with IEC
61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to
2.7 GHz
d
= 1.2
80 MHz
to 800 MHz
122
| Drystar 5500, Drystar 5503 | Remarks for HF-emission and immunity
2901J EN 20191018 0826
Содержание 5364/100
Страница 1: ...Drystar 5500 Drystar 5503 5364 100 5364 300 User Manual 2901J EN 20191018 0826 ...
Страница 12: ...12 Drystar 5500 Drystar 5503 Drystar 5500 Drystar 5503 2901J EN 20191018 0826 ...
Страница 58: ... Audio signals The keypad The display 58 Drystar 5500 Drystar 5503 Introduction 2901J EN 20191018 0826 ...
Страница 87: ...The menu structure Drystar 5500 Drystar 5503 Advanced Operation key operator mode 87 2901J EN 20191018 0826 ...
Страница 160: ...4 Open the upper lower input tray 160 Drystar 5500 Drystar 5503 Plug Play Installation Manual 2901J EN 20191018 0826 ...