Chapter 4 Maintenance Procedures
93
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
NOTE
Using a squeeze bottle with a right-angle elongated nozzle is a convenient way to fill the ultrasonic
wash.
6.
Close the main door
WARNING
!
WARNING!
Do not place hands in the main door pathway when closing it.
WARNING
!
WARNING!
When lowering the main door, do not let go of the handle until the door is completely
closed. The gas springs support the door as it is lowered, but stop providing support when the
door is almost closed. If you do not hold onto the door handle, the door may slam closed.
7.
Tap the
HOME
button in the
safety interlock is disengaged!
message.
Wait for the pin array assembly to move to home position (all the way left, forward,
and up).
8.
Tap
CLEAN DAILY
on the Maintenance screen
.
The pins are immersed in ethanol for 30 minutes. An on-screen message displays the
time remaining for the operation. Wait for the operation to complete.
NOTE
If you cancel the pin cleaning operation, the pins are put through a rinse-clean cycle before the
operation is canceled. Wait for the rinse-clean cycle to finish.
Figure 4.5 Ultrasonic Wash Deck Location
Ultrasonic wash