Chapter 2 Dispensing to SpectroCHIP Arrays
29
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
To check tank status:
1.
Tap the
STATUS
button on the
Main Menu
(or any other screen).
.
2.
On the
Machine Status
screen that appears, tap the
Page 2
tab (bottom of the
screen).
3.
Check the tank status on
Page 2
NOTE
NOTE
If you are using an external supply tank and the
supply tank level (low)
indicator is lit, you must fill
the internal supply tank to at least half-full. See
Checking the external supply and waste tanks,
4.
Tap the
EXIT
button
to return to the previous screen.
Performing pin priming
(optional)
In some cases, it may be necessary to "prime" the pins prior to dispensing. Priming the
pins wets the surface of the pins to enable better pickup of analyte solution during
dispensation. Pin priming is usually not required, but may improve spotting performance
in low relative humidity (rH) environments.
Figure 2.10 STATUS Button in the Main Menu
Figure 2.11 Machine Status, Page 2
A bright green light
means the supply
tank is low and must
be filled.
A bright green light
means the waste
tank is full and must
be emptied.