Chapter 2 Dispensing to SpectroCHIP Arrays
28
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
3.
To delete a user:
a. Tap the user name to be deleted
b. Tap the
DEL USER
button
in the
User Setup
screen.
c. Tap
YES
in the confirmation message that appears.
4.
To edit user login information (for example, change the password or user level):
a. Delete the existing user (follow step
).
b. Add a new user with the same login name, but new password and/or user level
).
2.4 Pre-run checks
Performing daily
maintenance
If you are the first person of the day to use the instrument, complete the daily
maintenance procedures. See
for daily maintenance instructions.
Checking the supply
and waste tanks
NOTE
NOTE
This section applies to the internal tanks. If you are using external tanks, see
supply and waste tanks, page 141
Before starting a dispensing run, confirm that the:
•
Supply tank is filled with Type 1, >18.2
M
Ω
water (a full tank is sufficient to process up
to ten 384 SpectroCHIP Arrays). See
supply tank.
•
Waste tank is empty (an empty tank has the capacity to process ten 384 SpectroCHIP
for details on draining the waste tank.
Tank status
If the
STATUS
button
(at the top of most screens) is blinking, it means that the supply
tank is low and/or the waste tank is full.
Figure 2.9 Select and Confirm the User to be Deleted