Appendix A Changing the Pin Format
118
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
A.1 Adapting an instrument to 6-pin or 1-pin format
This section explains the steps to install a 6- or 1-pin insert, and to remove an insert to
return to the standard 24-pin format.
1.
Go to the Maintenance screen by tapping the
Tools
button on the Main Menu.
If you are not on the Main Menu, tap the
BACK
button
on any screen.
2.
Tap
CHANGE INSERT
on the
Maintenance
.
3.
Follow the instructions that appear:
a. Open the main door.
b. Remove the vacuum cover and calibrant reservoir by simply lifting them off the dry
station. If the calibrant reservoir is a little difficult to remove, rock it slightly side-to-
side as you lift
.
Figure A.2 Go to the Maintenance Screen
Figure A.3 View the CHANGE INSERT Instructions