THANK YOU
On behalf of the entire team at Thane Direct, we appreciate
your recent purchase of Ageless WOnDer™.
Ageless Wonder™ delivers gentle microcurrent impulses to
minimize wrinkles and to restore and tone facial muscles for
a more youthful appearance.
• Improve facial tone
• Rejuvenate the skin
• Reduce puffiness
• Even out skin tone
Enjoy your Ageless WOnDer™
and, once again,
thank you!
TROUBLESHOOTING GUIDE
PROBLEM
POSSIBLE CAUSE
SOLUTION
lCD Display does
not light up at all
- Batteries not installed
- Batteries installed incorrectly or
battery tray not properly locked
into place
- Batteries are dead
- Install Batteries
- Check polarity “+” , “-”
- remove the battery tray and
re-insert it until it locks into place.
A confirmation beep should sound.
- replace Batteries
lCD Display lights
up, but no
sensation is felt.
- no skin contact
- Application wand is not
plugged in correctly
- Application wand is not
attached to headset correctly
- Microprocessor freeze
- Battery is low
- Check to see if the conductive surface is
covered with a layer of gel
- Check that wand plug is connected firmly to
the control unit
- Check operation with handheld wand
first to isolate the problem. If handheld
operation works, attach application wand to
the headset and-
- Check that both wands are properly
attached to the headset.
- Choose
mode and Increase intensity
by pushing the + button
- Perform a reseT, remove batteries for 1
minute for full reseT
- Change Batteries
lCD Display lights
up, but nothing
occurs.
- Intensity set to low level
- Conductive sponge is not wet
enough
- The wands are positioned
incorrectly
- Conductive sponge has oil
deposits.
- Battery is low
- Choose
mode and Increase intensity
by pushing the + button
- Put more gel on the conductive surface
- Move wand around to locate your trigger
points.
- Clean conductive surface.
- Change Batteries.
lCD Display lights
up, but none of
buttons work.
- Microprocessor freeze
- Battery is low
- Perform a reseT, remove batteries for 1
minute for full reseT
- Change Battery
IMPORTANT: POWER RESET
If the unit does not switch on, or if you can not change to different programs, the microprocessor may have
“frozen”. Please perform a simple
RESET
. remove batteries and replace them after 1 minute.
16