Performing a control solution test
The control solution is used to:
•
Make sure that the AgaMatrix
WaveSense JAZZ test strips work with
JAZZ™ DoseCoach
®
.
•
Practice measuring without having to
use your own blood.
Do
4
4
Control solution should only
be used for 90 days after first
opening the bottle or until the
date of expiry printed on the
label, whichever comes first.
4
4
You should only use AgaMatrix
control solution with JAZZ™
DoseCoach
®
.
4
4
Always put the cap on the
control solution bottle right
after you have used it.
4
4
Do not add water or any other
liquid to the control solution.
4
4
Control solution tests should
be performed within the
temperature range of 10°C to
40°C.
Control solution tests should be
done when you:
•
First get your JAZZ™ DoseCoach
®
.
•
Think your JAZZ™ DoseCoach
®
or test strips are not working
properly.
•
Think your readings are
inaccurate.
•
Have dropped or damaged your
JAZZ™ DoseCoach
®
or got liquids
such as water on it.
•
Are told by your healthcare
professional to do so.
Содержание JAZZ DoseCoach
Страница 1: ...Dose Helper and blood glucose meter OWNER S GUIDE ...
Страница 8: ...Contents 6 This page has been left blank on purpose ...
Страница 26: ...Section 1 Key information about JAZZ DoseCoach 24 This page has been left blank on purpose ...
Страница 108: ...Section 4 Using Dose Helper 106 This page has been left blank on purpose ...
Страница 119: ...Section 5 Reviewing your history data 117 This page has been left blank on purpose ...
Страница 176: ...Section 8 Questions and answers on JAZZ DoseCoach 174 This page has been left blank on purpose ...
Страница 227: ...Section 11 JAZZ DoseCoach specifications 225 This page has been left blank on purpose ...
Страница 228: ...8100 10395 Rev A ...