PRECAUTIONS
Notice for the LCD Display
Accessory equipment connected to the analog and digital interfaces must be in compliance with the
respective nationally harmoniz
e d IEC standards (i.e. IEC 60950 for data processing equipment, IEC 60065
for video equipment, IEC 61010-1 for laboratory equipment, and IEC 60601-1 for medical equipment).
Furthermore all configurations shall comply with the system standard IEC 60601-1. Everybody who
connects additional equipment to the signal input part or signal output part configures a medical system,
and is therefore, responsible that the system complies with the requirements of the system standard
IEC 60601-1. The unit is for exclusive interconnection with IEC 60601-1 certified equipment in the patient
environment and IEC 60
certified equipment outside of the patient environment. If in doubt, consult the
technical services department or your local representative.
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle
mark
e d “Hospital Only” or “Hospital Grade”.
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
The single device output analog signals through ADC element (Analog DigitalConvert) conversion to
become a digital signal and the video signal is via Video Decorder conversion. It has become the same
digital signal, these signals via Scaler IC as zoom in or out action and digital image processing, then
through the cable line transmission LVDS signals to one of the LCD module. The last by the clock
controller
(Timing Controller, TCON), the clock
signal is transmitted to the drive IC on the panel and turn on Back
l ight
for LCD module light source by Scaler control.
WARNING
– No protection against the ingress of water : IPX0
WARNING
– Do not modify this equipment without authoriz
a tion of the manufacturer.
Installation and OSD adjusting should only be carried by manufacturer trained and authoriz
e d personnel.
WARNING
– To avoid risk
of electric shock
, this equipment must only be connected to a supply mains with
protective earth.
CAUTION
: This adapter Manufacturer/model is a forming part of the medical device.
Power by class I power supply.
Adapter manufacturer/model
ADAPTER TECH: ATM065T-P240
Input/output: 100-240V~50-60Hz
, 24V(
) 2.71A.
WARNING
: Use suitable mounting apparatus to avoid risk
of injury.
WARNING
The equipment not suitable for use in the presence of a flammable anesthetic mixture with air
or with oxygen or nitrous: Not AP or APG Category
CAUTION
: No applied part.
Mak
e sure the user not to contact SIP/SOPs and the patient at the same time.
CAUTION
Transport should only be undertaken in a flat surface.