Mounting and commissioning
– English –
DIT 10
38
1
Cable gland (PG 9)
2
Mounting flange
Fig. 9: Mounting with cable gland (PG 9)
2.
Insert the enclosed cable gland (PG 9) and fasten it with the en-
closed nut.
3.
Insert the cable of the pressure sensor into the cable gland (PG
9) and fixate it at the correct length so that it is odour-tight.
6.8
Entering the tank data
Zero calibration has been performed.
Pressure sensor has been installed.
Tank shape
The arrow at the bottom of the display points to the unit "Litres".
The display shows the code of the selected tank shape. When
the unit is commissioned for the first time, the displayed code
is 0. Code 0 means that no tank shape has yet been selected.
1.
Use the keys
and
to set the code of the tank shape deter-
mined, see chapter 6.1, page 31.
2.
Press the key F, to confirm the setting. After that, you can enter
the tank volume.
Tank volume
The arrow at the bottom of the display points to the unit m³.
The display shows the tank volume set. The value 0000 means
that no tank volume has yet been entered.
1.
Use the keys
and
to enter the total volume of the tank fa-
cility determined. Press the key
to select the digit to be
changed. Then press the key
to change the value of the se-
lected digit.