39
Fig.E.2. Design of type B flow straightening vane.
3. Hole marking in straightening vanes is shown in Table E.1.
Table E.1. Coordinates of the holes in straightening vanes of A and B type
(D
– pipeline internal diameter)
Item
No.
X axis
Y axis
1
0
0
2
0
0.142
D
3
0
0.283
D
4
0
0.423
D
5
0.129
D
0.078
D
6
0.134
D
0.225
D
7
0.156
D
0.381
D
8
0.252
D
0
9
0.255
D
0.146
D
10
0.288
D
0.288
D
11
0.396
D
0
12
0.400
D
0.151
D
4. The straightening vane is installed in the pipeline at a distance of 1
2
pipeline’s DN from the last source of hydraulic resistance with respect
to the flow direction. When installing the straightening vane, the length
of a straight run before PEA is determined as the distance from the
source of hydraulic resistance to PEA.
re_buf311_eng_doc2.0
Содержание BUF
Страница 1: ...BATTERY POWERED ULTRASONIC FLOW METER AFLOWT BUF VERSION BUF 311 OPERATION MANUAL ISO 9001 2008 ...
Страница 2: ...2 Manufacturer quality management system is certified to ISO 9001 2008 URL http www aflowt com ...
Страница 31: ...31 APPENDIX B Wiring diagrams Fig B 1 BUF Flow meter wiring diagram ...