13

Testing Multimode or Single-mode Links

Step II - Verify Test Jumpers

9.  Disconnect the transmit jumper from the MFP. 
 

Do not disturb the transmit jumper at the OLS end!

10. If needed, change MFP adapter cap to match receive jumper connector.
 

Clean both ends of the receive jumper!

11. Connect receive jumper to the MFP.
12. Mate free ends of transmit and receive jumpers using the appropriate adapter. 
13. Verify that the insertion loss of this mated connector pair is under 0.75 dB - 

maximum allowed by TIA (AFL recommends 0.4 - 0.5 dB typical):

 

Observe the displayed value - insertion loss of the test jumpers in [dB]. 

 

If value is not acceptable, disconnect transmit and receive jumpers at the 
adapter, clean free ends of both test jumpers and repeat steps 13 & 14.

 

If value is still not acceptable, replace test jumpers and repeat steps 1-14.

14. If loss value is acceptable, disconnect transmit & receive jumpers at the adapter.
15. Move the MFP and OLS to opposite ends of the link to be tested.

OLS

 

MFP

Mandrel wrap - MM (shown), 
or 30 mm loop - SM

Transmit jumper

Do NOT 
disturb this 
connection

Adapter

Receive jumper

13

12

11

0.4 dB

Содержание MFI

Страница 1: ...MFIS Multi Fiber Identification System MFT Multi Fiber Tracer MFI Multi Fiber Identifier MFP Multi Fiber Power Meter Quick Reference Guide www AFLglobal com or 800 321 5298 603 528 7780 Test Inspection ...

Страница 2: ...r button is pressed to turn it Off 2 Signal Detect Indicators If valid signal detected both lights will flash and beeper will sound once every 3 seconds 3 No ID Indicator illuminates to indicate absence of a valid optical signal 4 Backlight Button When the MFI identifier is turned On press this button to toggle the Backlight feature On or Off 5 LCD Display The Battery icon appears on the MFI displ...

Страница 3: ...TP cable 2 Power Button Press this button to turn the MFT ON or OFF 3 Lasers Active Indicator illuminates to indicate that light source is turned On 4 Low Battery Indicator illuminates to indicate a low battery condition 1 2 3 4 MFT Multi Fiber Tracer 2 3 4 1 ...

Страница 4: ... power up until P on is displayed to disable the Auto Off feature Note MFP stays powered up until the Power button is pressed to turn it Off 2 Fiber ID Button enables disables Fiber ID mode When MFP operates in standard power meter mode pressing the Fiber ID button will enable Fiber ID mode that allows MFP to report loss and fiber ID from MFT s signal Pressing the Fiber ID button again will disabl...

Страница 5: ...er in µW 5 Ref Set Button works in both Fiber ID mode and standard power meter mode It provides two functions Press to display the stored reference level for the currently selected wavelength or multiple wavelengths Press and hold until HELD SET is displayed to store the currently measured level or multiple levels as the new reference levels Once the new reference is set an MFP switches to the dB ...

Страница 6: ... is designed to guide the fiber being tested to a precise position relative to the optical assembly The user should be careful to place fibers gently and in the provided fiber groove Forcing the fiber into the head assembly or misaligning it may induce optical losses above specifications Important The MFI will not falsely indicate Fiber ID Note Bright ambient room or outdoor light can cause the MF...

Страница 7: ...on Make sure the plunger is oriented correctly for the type of fiber to be tested If not lift the plunger from the two retaining pins Rotate the plunger such that the correct side is facing out and will be used for the fiber under test alignment Replace the plunger and cover The 250 900 RIB side faces out for the 250 µm and 900 µm coated and ribbon fibers testing 250 µm 900 µm Ribbon Fibers 250 90...

Страница 8: ... discriminate transmitted optical signals as follows If the fiber under test is active the MFI display will show fiber number if signal is detected Both arrows will alternately flash and an audible tone will beep every 3 seconds When no signal is present the MFI will illuminate the No Signal indicator and show no fiber on the display Additional Information for Ribbon Fibers Testing Ribbon fiber is...

Страница 9: ...n MFI will also display the ID of fiber under test Connect the MFT to the fiber network to be tested Turn On the MFT Turn On the MFI and insert the fiber to be tested in the fiber groove Pull down and hold the clamping trigger to depress the fiber under test into the optical assembly When the MFT signal is detected fiber ID will display on the MFI Verify if the fiber ID on display is matching fibe...

Страница 10: ... in the Fiber ID mode If MFP is in standad power meter mode press the Fiber ID button to switch to the Fiber ID mode Insert the fiber to be tested in the test port When the MFT signal is detected fiber ID will be displayed on the MFP along with fiber loss Verify if the fiber ID number on the MFP display matches the fiber number under test Also verify if the fiber loss is acceptable to test criteri...

Страница 11: ... type of this jumper must be the same as the fiber type normally connected to the output being measured 3 Mount the appropriate adapter cap on the MFP port This adapter cap must match the connector on the end of the test jumper you will connect to the MFP 4 Connect one end of the test jumper to the MFP adapter cap and the other end to the optical output to be measured 5 Press to select wavelength ...

Страница 12: ... ends of the transmit jumper 4 Connect the transmit jumper to the OLS output port MM or SM respectively 5 Mount adapter cap on the MFP must match free connector on transmit jumper 6 Connect transmit jumper free end to the MFP Display optical power in dBm 7 If measured power is outside of the normal range specified by manufacturer clean all fiber connections or replace the transmit jumper Repeat st...

Страница 13: ...r pair is under 0 75 dB maximum allowed by TIA AFL recommends 0 4 0 5 dB typical Observe the displayed value insertion loss of the test jumpers in dB If value is not acceptable disconnect transmit and receive jumpers at the adapter clean free ends of both test jumpers and repeat steps 13 14 If value is still not acceptable replace test jumpers and repeat steps 1 14 14 If loss value is acceptable d...

Страница 14: ... Multimode Link Insertion Loss 16 Connect free ends of transmit and receive jumpers to the link under test Clean jumper end that connects to patch panel prior to every test 17 MFP will measure and display the insertion loss of the link under test 18 Record link insertion loss at the current test wavelength 19 Repeat steps 17 19 for all links to be tested at the current wavelength ...

Страница 15: ... a battery icon appears at the top of the MFI display the 2 x 1 5 V AAA alkaline batteries require replacement To replace the discharged batteries Remove the retaining screw and slide the battery plate away from the unit Replace the discharged batteries Replace the battery plate and retaining screw MFT and MFP When the battery indicator lights on the MFT MFP unit the 2 x AA alkaline batteries requ...

Страница 16: ...AFL Test and Inspection products are warranted against defective material and workmanship for a period of 1 one year from the date of delivery to the end user Optional Extended Warranty starts at the end of the standard 1 one year warranty period Any product that is found defective within the warranty period will at the discretion of AFL be repaired or replaced Warranty will be voided if the produ...

Страница 17: ...Thank you for choosing AFL Test Inspection www AFLglobal com or 800 321 5298 603 528 7780 Test Inspection 2018 AFL all rights reserved MFIS QRG 1ENG Revision AA 2018 10 04 AFLTest Inspectio n ...

Отзывы: