AFi RT-913 Скачать руководство пользователя страница 6

 

6

CHANNEL BIAS/TERMINATION SWITCHES 

Switches are available internally that allow offset bias and termination features to be activated when 
using RS485 data. These switches also allow termination features to be activated when using RS422 
data. In order to reconfigure the RS485/422 channel, the rack card cover needs to be removed. To 
open the RT-913, remove the two screws at the left rear of unit near data and fiber connections. Also 
remove the two screws at the right front of unit (top of cover). The configuration switch banks are 
located at the lower right corner of PCB next to the data terminals (See drawing below). 
The RT-913 is shipped with these switches in the off 
(down) position. When transmitting RS232 or 
Manchester data, the bias and termination switches 
must remain in the off (down) position. The top (S1) 
switches correspond to channel 1. The middle and 
bottom switch banks correspond to channel 2 and 3 
respectively. The chart below describes the layout of 
the top (S1) switch bank. The remaining switch banks 
follow the same layout.       

 

 
 
 
 
 
 
 
 
 
Please remember when using offset bias switches 
that they must be used in pairs. If switch # 1 is up then 
switch # 3 must also be up. The same situation applies for switch # 4 and switch # 6. Also, if using 
offset bias, it is important that the corresponding termination resistor switch be in the on (up) position. 
Using offset bias without a termination on the line will cause communications to fail. An explanation 
follows on general bias and termination guidelines.  

 

RS485 DATA TERMINATION 

The RS485 protocol is an expanded version of the original RS422 protocol. RS485 differs from RS422 
in the ability of the transmitter devices to go into a high impedance (Hi-Z) state. This allows multiple 
transmitter devices to reside on the same wire pair. The software must dictate a protocol that allows 
only one device to transmit at any one time to prevent data crashes. In many cases the system head 
end controller will continuously poll data from all remote devices. The remote devices all respond back 
to the head end (one at a time) as they are addressed. The driver chips that are used in RS485 
communications are capable of changing into their high impedance state very rapidly. On even short 
lengths of wire there can exist a residual voltage after a driver circuit turns off. This can interfere with 
circuits that are used to detect the Hi-Z state. It is very important that the copper communications lines 
be terminated with resistors across the data wire pair. The best place to locate such resistors is at the 
furthest electrical devices at the ends of the wire pair. For instance, if several RS485 devices are 
connected in a daisy chain fashion, the wire connection would loop across all devices in a chain. The 
furthest two points in the chain would need to be terminated. 

 
OFFSET BIAS – RS485 

The RS485 specification requires receivers to detect input signals down to 200mVp-p of voltage level. 
In many cases this can cause systems to be sensitive to noise on the data wires. In an effort to 
eliminate the effects of low levels of noise, some manufacturers of equipment that communicate using 
RS485 have introduced a small voltage bias to the data lines. This is usually accomplished using a 470 
Ohm resistance to +5V on the positive line and 470 Ohm resistance to ground on the negative line. 
When used in conjunction with the appropriate termination resistors referred to in the previous section, 
this introduces about a 300 mV offset, improving noise immunity. 

Содержание RT-913

Страница 1: ...Copyright 2005 American Fibertek Inc 0824JD Instruction Manual RT 913 Video Transmitter With Bi directional Multi Protocol Data...

Страница 2: ...ion 4 Power Source 4 Power Connection 4 Fiber Connection 4 Video Input Connection 4 Data Input Output Connections 5 Typical System Data Connections 5 Data Configuration 5 RS485 Data Signals 6 Offset B...

Страница 3: ...CONFIGURATION section for changes to the default switch settings Channel 1 RS422 data Channel 2 RS485 4 wire or Manchester data Channel 3 RS232 data The RT 913 is designed to operate with the MR 913 o...

Страница 4: ...to the unit via a four finger backplane connector The RR 913 can be inserted into the subrack or removed from the subrack with power applied to the backplane FIBER CONNECTION The fiber optic connectio...

Страница 5: ...ta Out of the copper device is connected to the Data In of the RT 913 The plus terminal of the copper device is connected to the plus terminal of the RT 913 and the minus is connected to the minus The...

Страница 6: ...smitter devices to go into a high impedance Hi Z state This allows multiple transmitter devices to reside on the same wire pair The software must dictate a protocol that allows only one device to tran...

Страница 7: ...LED indicator is provided to monitor the data coming in from the fiber through the RT 913 and out onto the electrical interface The intensity of this indicator will vary with input data patterns howe...

Страница 8: ...Material Authorization RMA number is required to send the products back in case of return All returns must be shipped prepaid This warranty is void if the products have been tampered with This warran...

Отзывы: