60
microAeth
®
MA Series MA200, MA300, MA350
Operating Manual
microAeth
®
MA Series MA200, MA300, MA350
Operating Manual
61
AethLabs
AethLabs
Verbose Structure Details
Header
Units / Format
Example Data
Description
Accel Y
Integer
-8
Measured instantaneous acceleration in the y-axis.
Accel Z
Integer
-255
Measured instantaneous acceleration in the z-axis.
Tape position
Whole number
1
Value of tape position counter. Increments by 1 after
each tape advance. Counter is reset when 'Release
Tape' is used to release the tape cartridge no matter
the position of the newly inserted tape cartridge.
Flow setpoint
Milliliters per minute
(Whole number)
100
Target value of the instrument total flow rate in
milliliters per minute. Used in SingleSpot™ and
DualSpot® sampling modes.
Flow total
Milliliters per minute
(Decimal number)
100.10
Measured total flow through the instrument in
milliliters per minute. Measured in SingleSpot™ and
DualSpot® sampling modes.
Flow1
Milliliters per minute
(Decimal number)
60.23
Measured flow through sample spot sense1 of the
instrument in milliliters per minute. Measured in
SingleSpot™ and DualSpot® sampling modes.
Flow2
Milliliters per minute
(Decimal number)
39.87
Calculated flow through sample spot sense2 of the
instrument in milliliters per minute. Calculated in
DualSpot® sampling mode. No flow in SingleSpot™
sampling mode.
Sample temp
Degrees celsius
(Decimal number)
26.6705474853515
Measured temperature of instrument sampling
stream in degrees celsius.
Sample RH
Percent
(Decimal number)
22.6140441894531
Measured relative humidity of instrument sampling
stream in percent.
Sample
dewpoint
Degrees celsius
(Decimal number)
10.3270139694213
Calculated temperature of dewpoint of instrument
sampling stream in degrees celsius.
Int pressure
Pascal
(Natural number)
100692
Measured pressure inside instrument case in
pascals.
Int temp
Degrees celsius
(Decimal number)
35.375
Measured temperature inside instrument case in
Celsius.
Optical config
<SS or DS>-XX-X-
X-X-XX
DS-UV-B-G-R-IR
Identification of sampling mode (SingleSpot™ (SS)
or DualSpot® (DS)) and only the optical wavelengths
used for each measurement are displayed.
UV Sen1
Counts
(Whole number)
710434
Measured optical intensity of ultraviolet (UV)
wavelength for sample spot sense1 in counts.
Measured in SingleSpot™ and DualSpot® sampling
modes.
UV Sen2
Counts
(Whole number)
724526
Measured optical intensity of ultraviolet (UV)
wavelength for sample spot sense2 in counts.
Measured in DualSpot® sampling mode.
UV Ref
Counts
(Whole number)
907699
Measured optical intensity of ultraviolet (UV)
wavelength for reference spot in counts. Measured
in SingleSpot™ and DualSpot® sampling modes.
UV ATN1
Attenuation units
(Decimal number)
20.031862
Calculated attenuation of ultraviolet (UV)
wavelength for sample spot sense1 in attenuation
units. Calculated in SingleSpot™ and DualSpot®
sampling modes.
UV ATN2
Attenuation units
(Decimal number)
12.559727
Calculated attenuation of ultraviolet (UV)
wavelength for sample spot sense2 in attenuation
units. Calculated in DualSpot® sampling mode.
Verbose Structure Details
Header
Units / Format
Example Data
Description
UV K
Unitless
(Decimal number)
0.0046992627903819
Calculated loading compensation factor for
ultraviolet (UV) wavelength using sample spots
sense1 and sense2, used to compensate BC1 value
resulting in BCc. Calculated in DualSpot® sampling
mode.
Blue Sen1
Counts
(Whole number)
751124
Measured optical intensity of blue wavelength
for sample spot sense1 in counts. Measured in
SingleSpot™ and DualSpot® sampling modes.
Blue Sen2
Counts
(Whole number)
759335
Measured optical intensity for blue wavelength
for sample spot sense2 in counts. Measured in
DualSpot® sampling mode.
Blue Ref
Counts
(Whole number)
843783
Measured optical intensity of blue wavelength for
reference spot in counts. Measured in SingleSpot™
and DualSpot® sampling modes.
Blue ATN1
Attenuation units
(Decimal number)
15.778888
Calculated attenuation of blue wavelength for
sample spot sense1 in attenuation units. Calculated
in SingleSpot™ and DualSpot® sampling modes.
Blue ATN2
Attenuation units
(Decimal number)
9.550997
Calculated attenuation of blue wavelength for
sample spot sense2 in attenuation units. Calculated
in DualSpot® sampling mode.
Blue K
Unitless
(Decimal number)
0.00198470731265842
Calculated loading compensation factor for blue
wavelength using sample spots sense1 and sense2,
used to compensate BC1 value resulting in BCc.
Calculated in DualSpot® sampling mode.
Green Sen1
Counts
(Whole number)
810122
Measured optical intensity of green wavelength
for sample spot sense1 in counts. Measured in
SingleSpot™ and DualSpot® sampling modes.
Green Sen2
Counts
(Whole number)
829080
Measured optical intensity of green wavelength
for sample spot sense2 in counts. Measured in
DualSpot® sampling mode.
Green Ref
Counts
(Whole number)
902411
Measured optical intensity of green wavelength for
reference spot in counts. Measured in SingleSpot™
and DualSpot® sampling modes.
Green ATN1
Attenuation units
(Decimal number)
13.872042
Calculated attenuation of green wavelength for
sample spot sense1 in attenuation units. Calculated
in SingleSpot™ and DualSpot® sampling modes.
Green ATN2
Attenuation units
(Decimal number)
8.236315
Calculated attenuation of green wavelength for
sample spot sense2 in attenuation units. Calculated
in DualSpot® sampling mode.
Green K
Unitless
(Decimal number)
0.00123747368343174
Calculated loading compensation factor for green
wavelength using sample spots sense1 and sense2,
used to compensate BC1 value resulting in BCc.
Calculated in DualSpot® sampling mode.
Red Sen1
Counts
(Whole number)
794974
Measured optical intensity of red wavelength
for sample spot sense1 in counts. Measured in
SingleSpot™ and DualSpot® sampling modes.
Red Sen2
Counts
(Whole number)
759396
Measured optical intensity of red wavelength
for sample spot sense2 in counts. Measured in
DualSpot® sampling mode.
Red Ref
Counts
(Whole number)
888090
Measured optical intensity of red wavelength for
reference spot in counts. Measured in SingleSpot™
and DualSpot® sampling modes.