4.2 Diagnostics
The IQM has a number of sources for diagnostic information if a problem arises. These are described
below:
Event log
A large number of instrument events are logged on the event log file which is
located on the SD data card. This can be downloaded remotely (if a modem is
fitted) and in many cases can be used to determine whether a site visit is required.
Event Code Examples
Meaning
O3 :P.F. 2008/11/12 23:45
Pump failure
O3 module pump failure on 12 Nov
2008 at 23:45
NO2 :S.F. 2008/11/12 23:45
Sensor failure
NO2 module sensor failure on 12 Nov
2008 at 23:45
Power on: 2008/11/12 10:12
AQM Power turned on at 10:12 am on
12 November 2008
Zero Cal at 2008/11/12 11:24
Zero Calibration started at 11:24 on
12 November 2008
Config at 2008/11/12 13:12
0x01,05,30,40,50,B0,F8,00,00,00,00,00,00,00,00,00,
00
AQM configuration saved at 1:12 pm
on 12 November 2008
Sensor codes listed
O3 : NR 2008/11/12 12:42
O3 module no data report at 12:42 pm
on 12 November 2008
NO2 Scrubber FAILED: 2008/11/12 15:36
NO2 scrubber failed at 3:36 pm on 12
November 2008
Messages
This is located on the AQM PC software tool bar. If the AQM is connected to the computer then event
messages will be written to the Messages window in real-time.
Diagnostics
This is located in the AQM PC software. If the AQM is connected to the computer then individual sensor
modules can be interrogated to determine if there is a problem not picked up by normal fault detection as
well as fine tune the sensor performance. Only qualified personnel should use this menu as incorrect use
may result in sensor malfunction. Please consult Aeroqual Technical Support to understand how to use this
feature.