2.2.1 Wire AC mains power and initial start-up.
Remove the front cover and also remove the mains AC power box cover. Then feed the mains power
cable through the gland on the bottom floor of the AQM 65 as shown in Figure 2-2.
Online
reference
training.aeroqual.com AQM 65 technical training Section 2.3
Online video
https://vimeo.com/119735823
Figure 2-2 Remove the front cover to wire mains power to the AQM 65
Pull the 12 VDC fuse out from the fuse holder “off position” before testing the mains connection.
Figure 2-3 The 12V DC fuse can be pulled out to turn off the AQM or pushed in to power the AQM 65
When you apply mains AC power (and with the 12 DC fuse in the “off” position) you should hear the
internal fan (blower) start up and begin circulating air inside the box. Air will come out from the top
vent as shown in the image on the right in Figure 2-4
Nothing else should be on.
IMPORTANT!
If you wish to test 12 DC power to the rest of the AQM 65 now, then you should remove the power
connection to the PM electronics module, otherwise you could damage the optical engine. Pull the
power connection out from the PM pump module as shown in Figure 2-3 on the left.