28
Aeroneb
®
Professional Nebulizer System Instruction Manual
Cleaning, disinfection and
sterilization
This section describes how to clean, disinfect, sterilize and
inspect Aeroneb Pro system components. It is important that
Aeroneb Pro device components are cleaned and sterilized
prior to first patient use. The components are:
Nebulizer unit (including filler cap)
T-piece (including T-piece plug) for adult and
pediatric
Neonate adapters
Control module
†
, control module cable
†
, and
AC/DC adapter
†
Mounting bracket
†
†
Components not to be autoclaved.
CAUTION:
Always clean, sterilize and disinfect in accordance
with current hospital protocols.
To avoid damage to the nebulizer:
Autoclave according to specified directions and
temperature given in the “Cleaning, disinfection and
sterilization” section of this Instruction Manual only.
Any deviation from directions given in this Instruction
Manual may cause damage to the nebulizer unit and
render it inoperable.
Do not apply undue pressure to the domed aperture
plate in the center of the nebulizer.
Do not push out the Aerogen Vibronic
®
Aerosol
Generator
Содержание Aeroneb Pro
Страница 1: ...Instruction Manual ...
Страница 2: ... ...
Страница 3: ...Aeroneb Professional Nebulizer System Instruction Manual ...
Страница 4: ... This page has been intentionally left blank ...
Страница 48: ...42 Aeroneb Professional Nebulizer System Instruction Manual This page has been intentionally left blank ...
Страница 49: ...Aeroneb Professional Nebulizer System Instruction Manual 43 Appendix 1 EMC Tables ...
Страница 57: ......