OPERATION
3-1-48
Fig. 3-1-42 Sweep stopped
Fig. 3-1-43 RF level transferred
Aborting the sweep
Press
[Abort Sweep]
. The sweep is reset and the RF (or LF) signal is removed from the
appropriate output socket. The
Sweep Parameters Menu
as shown in Fig. 3-1-37 is displayed. If
the sweep is restarted with the signal disabled, the
[Abort Sweep]
key is not displayed thus
giving the user an indirect indication that no signal is being output from the instrument.
LOCAL
Stop
Sweep
Reset
Sweep
Transfer
Level Start:
Level Stop:
Sweep Status: PAUSED AT +6.4 dBm
Sweep Mode: INTERNAL SINGLE
Sweep Type: RF LEVEL
Number of:
Steps
Step Time:
dBm
dBm
ms
-144.0
+13.0
1000
-----
50
C1901
RF
Level
Carrier
Freq.
LOCAL
FM
ON/OFF
Select
Source
FM
Devn.
Source
Freq: F4
Low
Intermod.
Wideband
FM
AM
Φ
M
Carrier :
Freq.
RF Level :
dBm
Single Modulation Mode
Modulation ENABLED
MHz
Int Std: 10 MHz
FM:
kHz
Int F4:
1.0000 kHz
10.0
ON
626.407 6800
+6.4
ON
C1902