AC/DC Current Probe Model SL206 

 

- 6 - 

Specifications 

ELECTRICAL SPECIFICATIONS 

Current Range: 

10mA to 80A

DC

; 50mA to 60A

AC

 

Output Signal:

 1mV/mA and 10mV/A

AC/DC

 

Accuracy and Phase Shift*:

 

Range 1mV/mA 

(1V/A) 10mV/A 

Current Range

(1)

 

10mA to 2A

DC

 

10mA to 1.5A

AC

 

50mA to 80A 

Output Signal 

1V at 1A

AC/DC

 

1V at 100A

AC/DC

 

Accuracy % 

of Reading 

2% of Reading  

± 5mA 

50mA to 50A

DC

: 4%  

of R ± 20mA 

50mA to 40A

AC

: 4%  

of R ± 20mA 

50 to 80A

DC

: ±12% of R 

40 to 60A

AC

: ±12% of R 

Frequency Range 

DC to 2kHz 

DC to 8kHz 

Phase Shift 

DC to 65Hz: 1° 

DC to 65Hz: 1° 

Load Impedance 

100k

 min 

10k

 min 

Noise 

DC to 1Hz: 1.5mV 

1Hz to 10kHz: 14mV

10 to 100kHz: 18mV

DC to 1Hz: 15µV 

1Hz to 10kHz: 140µV 

10 to 100kHz: 180µV 

(1) The maximum permitted value is only limited by the cross section of the primary conductor. 

Saturation of the magnetic core of the clamp occurs, which is not dangerous. A high residual 
magnetism may remain which can be eliminated by demagnetizing the clamp. 

Overload:

 120A continuous 

Working Voltage:

 600Vrms max 

Common Mode Voltage:

 600Vrms max 

Battery: 

9V alkaline (NEDA 1604A, IEC 6LR61) 

Battery Life:

 68 hours typical 

Typical Consumption:

 6mA 

Low Battery:

 Green LED when 

6.5V 

(*Reference conditions: 23°C ± 5°K, 20 to 75% RH, 48 to 65Hz, external magnetic field < 40A/m, 
no external current carrying conductor, test sample centered, load impedance 1M

www.

GlobalTestSupply

.com

Quality AEMC Products Online at:

[email protected]

Содержание SL206

Страница 1: ...AC DC Current Probe Model SL206 USER MANUAL I 1 V A 1 0 m V A O F F Z E R O www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...

Страница 2: ...ning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer...

Страница 3: ...to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material YOU CAN NOW REGISTER...

Страница 4: ...rol and Connector Identification 5 Specifications 6 Electrical Specifications 6 Mechanical Specifications 7 Safety Specifications 8 Frequency Response Curves Typical 9 Operation 10 Connection 10 Zeroi...

Страница 5: ...conductors which may create noise Check the magnetic mating surfaces of the probe jaws these should be free of dirt rust or other foreign matter Do not use a probe that is cracked damaged or has defec...

Страница 6: ...ters Receiving Your Shipment Upon receiving your shipment make sure that the contents are consistent with the packing list Notify your distributor of any missing items If the equipment appears to be d...

Страница 7: ...it Hall sensor technology senses the magnetic field produced by the current measured and generates a millivolt DC or AC output signal The narrow jaw is designed for optimum use in crowded wiring in in...

Страница 8: ...0 to 100kHz 18mV DC to 1Hz 15 V 1Hz to 10kHz 140 V 10 to 100kHz 180 V 1 The maximum permitted value is only limited by the cross section of the primary conductor Saturation of the magnetic core of the...

Страница 9: ...condensation Maximum Cable Diameter 0 46 11 8mm Zero Adjustment 20 turn potentiometer 1 5A min Handle Lexan 920 A UL 94 V2 Case Protection IP20 per IEC 529 Drop Test 1 0 m on 38mm of oak on concrete t...

Страница 10: ...lution 2 Double insulation or reinforced insulation between primary or secondary and outer case of handle per IEC 1010 2 32 Electromagnetic Compatibility EN 50081 1 Class B EN 50082 2 Electrostatic di...

Страница 11: ...rent Probe Model SL206 9 Frequency Response Curves Typical 0 10 20 30 40 50 60 1 10 100 1000 10000 Hz 1 1 2 2 1 mV mA 10 mV A www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestS...

Страница 12: ...of residual magnetism often remains in the core This residual magnetism will cause the voltmeter to show a small DC reading even though there is no current passing through the jaws of the probe The re...

Страница 13: ...cal shock do not attempt to perform any servicing unless you are qualified to do so Avoid electrical shock and or damage to the instrument do not get water or other foreign agents into the electronic...

Страница 14: ...rded calibration data Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 repair aemc com Or co...

Страница 15: ...www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...

Страница 16: ...Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 www aemc com 99 MAN 100029 v12 07 11 www GlobalTestSupply com Quality AEMC Products Online at sales GlobalTestSupply com...

Отзывы: