MAINTENANCE

5.1 

Battery Installation

Under normal conditions, the battery will last up to a year of continuous 

recording unless the logger is restarted very frequently.
In the OFF mode, the logger puts almost no load on the battery. Use the 

OFF mode when the logger is not in use. Replace the battery once a year 

in normal use.
If the logger will be used at temperatures below 32°F (0°C) or is frequently 

turned on and off, replace the battery every six to nine months.

1. Make sure your logger is turned off (no blinking light) and all inputs

are disconnected.

2. Turn the logger upside down. Remove the four Phillips head screws

from the base plate, then take off the base plate.

3. Locate the two-wire (red/black) battery connector and attach the 9V

battery to it. Make sure that you observe polarity by lining up the

battery posts to the proper terminals on the connector.

4. Once the connector is plugged onto the battery, insert the battery

into the holding clip on the circuit board.

5. If the unit is not in record mode after installing the new battery, dis-

connect it and press the button twice then reinstall the battery.

6. Reattach the base plate using the four screws removed in Step 2.

Your logger is now recording (LED blinking). Press the 

PRESS

 button for 

five seconds to stop the instrument.

NOTE:

 For long-term storage, remove the battery to prevent dis-

charge effects.

5.2 

Cleaning

The  body  of  the  logger  should  be  cleaned  with  a  cloth  moistened  with 

soapy water. Rinse with a cloth moistened with clean water. Do not use 

solvent.

Содержание Simple Logger L320

Страница 1: ...SIMPLE LOGGER DC MODULE L320 L410 L430 E N G L I S H User Manual...

Страница 2: ...se record this number and purchase date for your records SIMPLE LOGGER DC MODULE MODEL _ _______________________________________________ SERIAL _ _______________________________________________ PURCHA...

Страница 3: ...ical Specifications 8 3 2 Mechanical Specifications 9 3 3 Environmental Specifications 9 3 4 Safety Specifications 9 4 OPERATION 10 4 1 Software Installation 10 4 2 Recording Data 11 4 3 Using the Sof...

Страница 4: ...APPENDIX B 20 Time Extension Recording TXR 20 Repair and Calibration 22 Technical and Sales Assistance 22 Limited Warranty 23 Warranty Repairs 23...

Страница 5: ...a scale for the model in use connect the logger and a Scale menu will appear Click on Scale and the scaling window for the model used will appear From the scaling window you may set a custom scale or...

Страница 6: ...ALWAYS connect the leads to the logger before inserting the leads to the test voltage ALWAYS inspect the instrument and leads prior to use Replace any defective parts immediately NEVER use the Simple...

Страница 7: ...istributor of any missing items If the equip ment appears to be damaged file a claim immediately with the carrier and notify your distributor at once giving a detailed description of any damage Save t...

Страница 8: ...ng Logging Two Blinks 2 Press to Stop Logging Stand By One Blink 3 Press Hold 5 sec to Shut Off Off No Blinks Warning All Data will be Deleted 4 1 Start Stop Button 2 Input Terminal Strip 3 Red LED In...

Страница 9: ...dition No Blinks OFF mode 2 2 Inputs and Outputs The left side of the Simple Logger incorporates 4mm safety banana jacks for the Models L410 and L430 and a screw connector for the Model L320 The right...

Страница 10: ...min resolution Scale Range Maximum Input Resolution 100 25 5mA 0 1mA 50 12 75mA 0 05mA 25 6 375mA 0 025mA 12 5 3 1875mA 0 0125mA L410 8 Bit 50 V min resolution Scale Range Maximum Input Resolution 100...

Страница 11: ...via DB9 connector 1200 Baud 3 2 Mechanical Specifications Size 2 7 8 x 2 5 16 x 1 5 8 73 x 59 x 41mm Weight with battery 5 oz 140g Mounting Base plate mounting holes or Velcro pads Case Material Polys...

Страница 12: ...h stored file One 9 pin serial port one parallel port for printer support CD ROM drive 1 Insert the Simple Logger CD into your CD ROM drive If auto run is enabled the Setup program will start automati...

Страница 13: ...Version 6 xx in the Options section 4 Click the Install button and follow the on screen prompts to install the software 4 2 Recording Data Connect the logger to the circuit to be tested NOTE Be sure t...

Страница 14: ...n the graph or in the tabular listing directly in the units appropriate to the measurement rather than mathematically converting a voltage or a current to the proper scale and value after the graph is...

Страница 15: ...lates to be used at a later date with newly downloaded loggers Creating and storing the scales and units here will save you time later on especially for frequently used scale settings Two buttons are...

Страница 16: ...programming window as seen on the screen 4 3 3 Creating Scales for Connected Loggers Connect the Simple Logger to the computer s serial port for download ing See main manual for downloading instructi...

Страница 17: ...25 4 would be a valid 5 character number Once the scale is defined click in the Unit box to program the engineering units to be displayed on the graph Up to 5 alphanumeric characters may be typed in...

Страница 18: ...late then take off the base plate 3 Locate the two wire red black battery connector and attach the 9V battery to it Make sure that you observe polarity by lining up the battery posts to the proper ter...

Страница 19: ...click on the desired file to open the Text Import Wizard 5 Review the selections in the first wizard screen and make sure that the following choices are selected Original Data Type Delimited Start Imp...

Страница 20: ...rom the main menu and select Columns from the drop down menu 2 Next click on column A at the top of the column to select the data then click on Edit from the main menu and select Copy to copy the enti...

Страница 21: ...d click on OK to format the column Table A 2 shows a typical spreadsheet with date time and value displayed It may be necessary to change the column width to see all the data A B C 12 02 04 11 45 AM 1...

Страница 22: ...er the completion of two hours of recording the logger continues recording by selectively overwriting previously stored data The Simple Logger also halves its sample rate to 2048 hr 1 76 seconds per p...

Страница 23: ...7 1 3604 48 8192 341 333 Table B 2 Sample Rate vs Recording Time Recording continues in this manner until the battery wears out or record ing is stopped For convenience in data analysis the recording...

Страница 24: ...ncludes calibration certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax...

Страница 25: ...period you may return the in strument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the...

Страница 26: ...24 Simple Logger Models L320 L410 L430 Notes...

Страница 27: ......

Страница 28: ...08 18 99 MAN 100212 v14 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: