10
AC/DC Current Probe Models MR461 and MR561
Noise:
60A Range:
DC to 1kHz
≤
8mV; DC to 5kHz:
≤
12mV; 0.1Hz to 5kHz:
≤
2mV
600A Range:
DC to 1kHz
≤
1mV; DC to 5kHz:
≤
1.5mV; 0.1Hz to 5kHz:
≤
500µV
Rise and Fall Time:
60A Range:
≤
100µs from 10 to 90% Vout
600A Range:
≤
70µs from 10 to 90% Vout
3.2.2 MECHANICAL
Humidity Influence:
10 to 90% RH @ reference temperature
≤
0.5%
Maximum Cable Diameter:
One 1.18” (30mm) or two 0.95” (24mm) or
two bus bars 1.24 x 0.39” (31.5 x 10mm)
Dimensions (without zero knob):
8.8 x 3.82 x 1.73” (224 x 97 x 44mm)
Weight:
15oz (440g)
3.3
MR561 Specifications
3.3.1 ELECTRICAL
Measurement Range:
150A Range:
0.2 to 100A
AC
(150A Peak); 0.4 to 150A
DC
1500A Range:
0.5 to 1000A
AC
(1400A Peak); 0.5 to 1500A
DC
Output Signal:
10mV/A on 150A range; 1mV/A on 1500A range
Accuracy:
150A Range:
0.5 to 100A: 1.5% reading ±0.5A;
100 to 150A
DC
only: 2.5% reading
1500A Range:
1 to 100A: 1.5% reading ±1A; 100 to 800A: 2.5% reading ±0.5A
800 to 1000A: 4% reading; 1000 to 1200A
DC
only: 4% reading