background image

 

ELECTRICAL SPECIFICATIONS

 

Nominal Range: 

10A

 

Measurement Range: 

0.1 to 12A 

Transformation Ratio: 

Voltage output

 

Output Signal: 

100mV/A (1V at 10A) 

Accuracy and Phase Shift*: 

Accuracy: 

0.1 to 0.5A: 

 2.5% Reading ± 0.1mV 

0.5 to 1A: 

 2.5% Reading 

1 to 12A:  

 1% Reading 

 

Phase Shift: 

.01 to 0.1A:  Not Specified 
0.1 to 1A: 

 5° 

1 to 5A: 

 3° 

5 to 12A:  

 2.5° 

*Reference conditions: 20°C to 26°C, 20 to 75% RH, 
external magnetic field <40A/m, 48 to 65Hz sine wave, 
distortion factor less than 1%, no DC component, no 
external current carrying conductor, test sample centered. 
Load impedance > 1M

Frequency Range: 

40 to 10kHz

 

Limit Operating Conditions: 

20A permanently to 10kHz; 
Derating above 10kHz: 20A x 0.1F (in kHz) 

Crest Factor: 

3 @ 10Arms with an error (due to CF) of 3% 

Working Voltage: 

600Vrms 

Common Mode Voltage: 

600Vrms 

Influence of Adjacent Conductor: 

< 15mA/A at 50Hz 

Influence of DC on AC signal: 

DC Current < 2A, < 5% 

Influence of Conductor Position in Jaw:  

0.5% of mV output @ 50/60Hz 

Influence of Frequency:  

40Hz to 1kHz: 0.5% of mV output 
1kHz to 10kHz: 2% of mV output 

Influence of Temperature: 

 150ppm/°K, or 0.15% of mV output per 10°K 

Influence of Humidity (10 - 90% RH): 

 0.2% of mV output per 10°K @ 18° to 28°C 

MECHANICAL SPECIFICATIONS 

Operating Temperature: 

14° to 131°F (-10° to +55°C)  

Storage Temperature: 

-40° to 158°F (-40° to +70°C) 

Operating Relative Humidity: 

10° to 35°C 85% RH 
(without roll-off above 35°C) 

Maximum Cable Diameter: 

One Ø 0.78" (20mm), bus bar 20 x 5mm 

Case Protection: 

IP 40 (IEC 529) 

Drop Test: 

Test per IEC 68-2-32: 
1.0m drop on 38mm of Oak on concrete. 

Mechanical Shock: 

Test per IEC 68-2-27 

Vibration: 

Test per IEC 68-2-6 

Dimensions: 

5.47 x 2.00 x 1.18" (139 x 51 x 30mm) 

Weight: 

180g (6.5oz) 

Polycarbonate Material: 

Jaws:  Polycarbonate with 10% fiberglass charge, 
 

Red UL 94 V0 

Case: Polycarbonate 920A Gray 

Opening Operations - Life: 

> 50,000 

Output: 

Double/reinforced insulated 5 ft (1.5m) lead with 
safety 4mm banana plug  

SAFETY SPECIFICATIONS 

      

      

     

Electrical: 

Double insulation or reinforced insulation between 
the primary or secondary and the outer case of the 
handle conforms to IEC 1010-2-32. 

Dielectric Strength: 

IEC 1010-2-32, Category III, Pollution Degree 2 
5550V, 50/60 Hz between primary, secondary  
and the outer case of the handle 

Electromagnetic Compatibility: 

EN 50081-1 Class B 
EN 50082-2 Electrostatic discharge 
   IEC 1000-4-2 
Radiated field IEC 1000-4-3 
Fast transients IEC 1000-4-4 
Magnetic field at 50/60 Hz IEC 1000-4-8 

ORDERING INFORMATION 

AC Current Probe MN307 ................. Cat #2116.23 

Accessories: 

Banana plug adapter 
(to non-recessed plug)  ....................... 

Cat #1017.45

 

E192383

2.56"

 (65 mm)

5.47"

 (139 mm)

.83"

 (21 mm)

Black

Red

.73"

2.00"

 (51 mm)

(18.5 mm)

.1.18"     

(30 mm)

Tactical

Barrier

(keep hands below

this barrier)

MN306 MN307 

Отзывы: