background image

ELECTRICAL SPECIFICATIONS

 

Current Range: 

100mA to 30A AC continuous cycle 
1 to 300A AC continuous cycle 
1 to 2000A AC continuous cycle for the  
full temperature range 
(2800 A if temperature is < 35°C or 95°F) 

Output Signal: 

100mV AC/A AC (3V at 30A) 90A peak 
10mV AC/A AC (3V at 300A) 900A peak 
1mV AC/A AC (3V at 3000A) 9000A peak 

Accuracy and Phase Shift*: 

30A RANGE 

Primary current 

1.5A 

6A 

30A 

Accuracy % 

2% ± 10mV (0.5A) 

Phase shift 

15° 

7° 

5° 

Probe Impedance: 

1000

 

300A RANGE 

Primary current 

15A 

60A 

300A 

Accuracy % 

2% ± 2mV (0.2mA) 

Phase shift 

3° 

1.5° 

1° 

Probe Impedance: 

100

 

 3000A RANGE 

Primary current 

150A 

600A 

3000A 

Accuracy % 

2% ± 0.5 mV 

Phase shift 

1.5° 

1° 

0.5° 

Probe Impedance: 

10

 

 (*Reference conditions: 23°C±3°K, 20 to 85% 
RH, 48 to 65 Hz, external magnetic field < 40 
A/m, no DC component, no external current 
carrying conductor, test sample centered.) 
Load impedance 

 1 M

Overload:

3200 A during 5 mn

 

Accuracy: 

Per IEC 185-26-27 

Ampere Second Product: 

100 A.s 

dV/Dt:

100 mV AC/A AC: dV/Dt = 400 mV/µs

 

 10 mV AC/A AC: dV/Dt = 50 mV/µs 
 1 mV AC/A AC: dV/Dt - 5 mV/µs 

Frequency Range: 

30Hz to 5kHz; current 

derating above 200Hz on the 3000A range

Load Impedance: 

> 100k

 

Working Voltage: 

600V AC 

Common Mode Voltage: 

600V AC

Open Secondary Voltage: 

10V Max.

 

Influence of Adjacent Conductor: 

0.005A/A AC 

Influence of Conductor in Jaw Opening:  

1.5% of reading 

Influence of Frequency from 30Hz to 5kHz:  

± 6% on all ranges 

Influence of DC Current: 

0.05% per A DC 

MECHANICAL SPECIFICATIONS

 

Operating Temp.: 

14° to 122°F (-10° to 50°C) 

Storage Temp.: 

-40° to 185°F (-40° to 85°C) 

Influence of Temperature: 

< 0.1% per 10°K 

Jaw Opening: 

3.54" (90mm) 

Maximum Conductor Size: 

Cable: 2.52" 

 max. (64mm) 

Bus bar: 1.97 x 5.31" (50 x 135mm)   
              2.52 x 3.94" (64 x 100mm) 

Envelope Protection: 

IP 20 (IEC 529)

Drop Test: 

500mm (IEC 68-2-32) 

Mechanical Shock: 

100 g (IEC 68-2-27) 

Vibration: 

10/55/10Hz, 0.15 mm (IEC 68-2-6) 

Polycarbonate Material: 

Handles:

 

10% fiberglass charged 

   polycarbonate UL 94 V0 
Jaws: ABS UL 94 V2 

Dimensions:  

4.72 x 12.40 x 1.89" (120 x 315 x 48 mm) 

Weight: 

2.65 lbs. (1200 g) 

Colors: 

Dark gray handles with red jaws 

Output: 

Two standard safety banana jacks (4 mm) 

SAFETY SPECIFICATIONS 

Electrical: 

Double insulation or reinforced insulation between 
primary or secondary and outer case of handle 
upon IEC 1010-2-32 

Common Mode Voltage: 

600V Category III, Pollution: 2 

ORDERING INFORMATION

 

Current Probe JM875 ........................ Cat  #2110.89 

Accessories:

Leads, set of 2, 5 ft. safety (1000V) .... 

Cat #2111.29 

Lead No. 211 for Probe 5 A Series ..... 

Cat #2950.06 

4.72"

(120 mm)

1.89"

(48 mm)

12.40"

(315 mm)

P1

P2

Ø=2.52" (64 mm)

Cable max.

1.97 x 5.31" (50 x 135 mm) or

2.52 x 3.94" (64 x 100 mm)

Bus Bar max.

Red

Black

S2

S1

Содержание JM875

Страница 1: ...hom it was purchased This warranty is void if the unit has been tampered with abused or if the defect is related to service not performed by AEMC Instruments Full warranty coverage and product registration is available on our website at www aemc com warranty html Please print the online Warranty Coverage Information for your records AC Current Probe Model JM875 User Manual DESCRIPTION The JM875 Ca...

Страница 2: ...n Secondary Voltage 10V Max Influence of Adjacent Conductor 0 005A A AC Influence of Conductor in Jaw Opening 1 5 of reading Influence of Frequency from 30Hz to 5kHz 6 on all ranges Influence of DC Current 0 05 per A DC MECHANICAL SPECIFICATIONS Operating Temp 14 to 122 F 10 to 50 C Storage Temp 40 to 185 F 40 to 85 C Influence of Temperature 0 1 per 10 K Jaw Opening 3 54 90mm Maximum Conductor Si...

Страница 3: ...n Secondary Voltage 10V Max Influence of Adjacent Conductor 0 005A A AC Influence of Conductor in Jaw Opening 1 5 of reading Influence of Frequency from 30Hz to 5kHz 6 on all ranges Influence of DC Current 0 05 per A DC MECHANICAL SPECIFICATIONS Operating Temp 14 to 122 F 10 to 50 C Storage Temp 40 to 185 F 40 to 85 C Influence of Temperature 0 1 per 10 K Jaw Opening 3 54 90mm Maximum Conductor Si...

Страница 4: ...sed or if the defect is related to service not performed by AEMC Instruments Full warranty coverage and product registration is available on our website at www aemc com warranty html Please print the online Warranty Coverage Information for your records AC Current Probe Model JM875 User Manual DESCRI PTION The JM875 Catalog 2110 89 is desi gned for use in industrial environments The squared jaws p...

Отзывы: