T1/T2 - 

T Button: 

  Pressing the T1/T2 - 

T Button selects in sequence T1, T2 and 

T. 

T is the 

temperature difference T2- T1. 

  The actual measurement T1, T2, T2-T1 is displayed in the upper left hand corner of the 

LCD. 

 

OPERATING INSTRUCTIONS 

OPERATION 

1.  Connect the K-type thermocouples to the instrument input. 

2.  Put the thermocouples near or on the sample tested. 

3.  Slide the function switch to the appropriate range or, if in doubt, select the highest 

setting (1°). If the measured temperature is below 200

, the higher resolution range 0.1

 

can be selected, if needed. 

4.  The Model CA863 may be in degrees Celsius (°C) at power-up. Press the back-light 

button 

 for more than 2 seconds to toggle over into degrees Fahrenheit (°F), if 

desired. The selected setting will remain for the next use.  

5. Thermocouples need a certain time to respond. Take the reading when the 

measurement has stabilized. 

6.  The HOLD button will freeze the measurement in the display, and will not update until 

released. 

7.  When finished, remove the thermocouple from the sample, unplug the thermocouple 

from the meter and slide the switch to OFF. 

 

SPECIFICATIONS 

ELECTRICAL 

Temperature Scale:

 

Celsius (°C) or Fahrenheit (°F) user-selectable 

Measurement Range (Dual Input): 

-58°F to 1999°F (-50°C to 1300°C) 

Resolution:

 

0.1°F/°C or 1°F/°C 

Accuracy (meter):

 

-58°F to 1999°F: ± 0.3% rdg + 2°F*  
-50°C to 1000°C: ± 0.3% rdg + 1°C* 
1000°C to 1300°C: ± 0.5% rdg + 1°C* 
*plus thermocouple 

Temperature Coefficient: 

0.1 times the applicable accuracy specification 
per °C from 0°C to 18°C and 28°C to 50°C. 

Input Protection: 

24VDC or 24Vrms maximum input voltage on any combination of inputs. 

Input Connector: 

Accepts standard miniature K-type thermocouple connectors. 

Temperature Response: 

Temperature indication follows Reference Temperature/Voltage Tables N.I.S.T. 
Monograph 175 Revised to ITS-90 for K-type thermocouples. 

 

GENERAL 

Display:

 

3½ digit liquid crystal display (LCD) with maximum reading of 1999

 

Over-Range

: “OL” is displayed 

Low Battery Indication:

 

The 

 is displayed when the battery voltage drops below the required level 

Sample Rate: 

2.5 times per second, nominal 

Operating Temperature: 

32° to 122°F (0°C to 50°C) at < 80%RH 

Storage Temperature: 

-4°F to 140°F (-20°C to 60°C), 0 to 80%RH with battery removed 

Altitude:

 2000m max 

Battery:

 

Standard 9V battery (NEDA 1604, IEC 6F22 006P, or equivalent) 

Battery Life:

 

200 hours typical with carbon zinc battery 

Dimensions:

 

6.81"(H) x 2.38"(W) x 1.5"(D) (173 x 60.5 x 38mm) 

Weight:

 

Approx. 6.4oz (181g) including battery 

Probe: 

The Thermometer CA863 is supplied with two 4 ft K-type thermocouples bead 

probe (Teflon sheath). Maximum sheath temperature 260°C (500°F). Probe accuracy 
±2.2°C (4

F) or ±0.75% of reading (whichever is greater) from 0°C to 800°C. 

Contact AEMC

®

 for other thermocouples available. 

 

SAFETY 

 

EN61010-1 (1995-A2), Protection Class III 
Overvoltage Category (CAT III, 24V), Pollution Degree 2 
Indoor Use 

 

 

www.