T1/T2 -
T Button:
Pressing the T1/T2 -
T Button selects in sequence T1, T2 and
T.
T is the
temperature difference T2- T1.
The actual measurement T1, T2, T2-T1 is displayed in the upper left hand corner of the
LCD.
OPERATING INSTRUCTIONS
OPERATION
1. Connect the K-type thermocouples to the instrument input.
2. Put the thermocouples near or on the sample tested.
3. Slide the function switch to the appropriate range or, if in doubt, select the highest
setting (1°). If the measured temperature is below 200
, the higher resolution range 0.1
can be selected, if needed.
4. The Model CA863 may be in degrees Celsius (°C) at power-up. Press the back-light
button
for more than 2 seconds to toggle over into degrees Fahrenheit (°F), if
desired. The selected setting will remain for the next use.
5. Thermocouples need a certain time to respond. Take the reading when the
measurement has stabilized.
6. The HOLD button will freeze the measurement in the display, and will not update until
released.
7. When finished, remove the thermocouple from the sample, unplug the thermocouple
from the meter and slide the switch to OFF.
SPECIFICATIONS
ELECTRICAL
Temperature Scale:
Celsius (°C) or Fahrenheit (°F) user-selectable
Measurement Range (Dual Input):
-58°F to 1999°F (-50°C to 1300°C)
Resolution:
0.1°F/°C or 1°F/°C
Accuracy (meter):
-58°F to 1999°F: ± 0.3% rdg + 2°F*
-50°C to 1000°C: ± 0.3% rdg + 1°C*
1000°C to 1300°C: ± 0.5% rdg + 1°C*
*plus thermocouple
Temperature Coefficient:
0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C.
Input Protection:
24VDC or 24Vrms maximum input voltage on any combination of inputs.
Input Connector:
Accepts standard miniature K-type thermocouple connectors.
Temperature Response:
Temperature indication follows Reference Temperature/Voltage Tables N.I.S.T.
Monograph 175 Revised to ITS-90 for K-type thermocouples.
GENERAL
Display:
3½ digit liquid crystal display (LCD) with maximum reading of 1999
Over-Range
: “OL” is displayed
Low Battery Indication:
The
is displayed when the battery voltage drops below the required level
Sample Rate:
2.5 times per second, nominal
Operating Temperature:
32° to 122°F (0°C to 50°C) at < 80%RH
Storage Temperature:
-4°F to 140°F (-20°C to 60°C), 0 to 80%RH with battery removed
Altitude:
2000m max
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22 006P, or equivalent)
Battery Life:
200 hours typical with carbon zinc battery
Dimensions:
6.81"(H) x 2.38"(W) x 1.5"(D) (173 x 60.5 x 38mm)
Weight:
Approx. 6.4oz (181g) including battery
Probe:
The Thermometer CA863 is supplied with two 4 ft K-type thermocouples bead
probe (Teflon sheath). Maximum sheath temperature 260°C (500°F). Probe accuracy
±2.2°C (4
F) or ±0.75% of reading (whichever is greater) from 0°C to 800°C.
Contact AEMC
®
for other thermocouples available.
SAFETY
EN61010-1 (1995-A2), Protection Class III
Overvoltage Category (CAT III, 24V), Pollution Degree 2
Indoor Use
www.
ShopAEMC
.ca
Shop for AEMC products online at:
1.877.766.5412