background image

 

99-MAN 100242 v2 5/03 

 

 OPERATOR MAINTENANCE 

 

WARNING:

 

To avoid possible electrical shock, disconnect the thermocouple connectors from the 
thermometer before opening the back cover. 

Use only factory specified spare parts for maintenance. The manufacturer will not be held 
responsible for any accident, incident, or malfunction following a repair done other than by 
its Service Center or by an approved repair center. 

 

Battery Replacement: 

A standard 9-volt battery supplies power. (NEDA 1604, IEC 6F22). 
The 

 appears on the LCD display when replacement is needed. 

To replace the battery: 

  Turn the Thermometer OFF 

  Remove the rubber holster. 

  Remove the screw from the back of the meter and lift off the battery cover. 

  Replace the battery, rear cover and holster. 

 

Cleaning: 

  Periodically wipe the case with a damp cloth and very mild soap, if needed. 

  Do not use any abrasives or solvents. 

  Do not let any liquid enter the case or sensor area. 

 REPAIR AND CALIBRATION 

To guarantee that your instrument complies with the factory specifications, we recommend 
that the unit be submitted to our factory service center at one-year intervals for 
recalibration, or as required by other standards. For repair and/or calibration: 

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

15 Faraday Dr. • Dover, NH 03820, USA 
(800) 945-2362 • (603) 749-6434 • Fax: (603) 742-2346 
www.aemc.com 

(Or contact your authorized distributor.) Costs for repair, standard calibration, and 
calibration traceable to N.I.S.T. are available. 

All customers must call for a return authorization number before returning any 
instrument. 

TECHNICAL AND SALES ASSISTANCE 

If you are experiencing any technical problems, or require any assistance, contact us at: 
(800) 343-1391 • (508) 698-2115 • Fax: (508) 698-2118 
www.aemc.com 

WARRANTY INFORMATION 

The Digital Thermometer Model CA863 is warranted to the owner for a period of 1 year 
from the date of original purchase against defects in manufacture. This limited warranty is 
given by AEMC

®

 Instruments, not by the distributor from whom it was purchased. This 

warranty is void if the instrument has been tampered with, abused or if the defect is related 
to service not performed by AEMC

®

 Instruments. 

For full warranty coverage detail and registration, go to www.aemc.com. 

To return the product for warranty repair or replacement free of charge, contact customer 
service at (800) 945-2362 or (603) 749-6434 to obtain a return authorization number. 

K-THERMOMETER 

(K-type thermocouple - Dual input) 

MODEL CA863 

USER MANUAL  

The AEMC

®

 Model CA863 is a portable, easy-to-use, compact-sized, dual input digital 

thermometer, designed for simple one-hand operation. It uses two external K-type 
thermocouples with standard miniature connectors as sensors. The thermometer is 
supplied with a protective holster and two K-type thermocouples. The Model CA863 also 
features a back-light LCD display, a data HOLD function, and a MAX recording function.  

 

 WARNING 

 

  Thermometers are not designed to work on live electrical samples. 

  Do not use this instrument when voltages at the sample point exceed 

24V

DC

 or 24Vrms. 

  Do not make temperature measurement in microwave ovens. Leaking 

microwaves may result in burns and serious injuries. 

  Repeated sharp flexing can cause the thermocouple leads to break. To 

prolong sensor life, avoid sharp bends in the thermocouple, especially 
near the connector. 

 

This symbol signifies CAUTION! and requests that the user refer to the user 
manual before using the instrument. 

PUSH BUTTONS 

Center (Yellow) Function Switch: 

Slide the center button to turn ON the Thermometer and select the 0.1

°

 or 1

°

 setting. Slide 

the button to OFF after use. 

Back-Light 

 and °C/°F Button: 

 Press 

the 

 

button to turn the Back-Light ON. 

 Press 

the   button again to turn the Back-Light OFF. 

 Press 

the 

 

button for 2 seconds to change the temperature scale to °C or °F. The 

temperature scale °C or °F is displayed in the lower right hand side corner of the 
display. 

HOLD/MAX Button: 

  Pressing the HOLD/MAX button once will initiate the data HOLD mode. The "HOLD" 

annunciator appears in the upper right hand corner of the display.  

  The Hold button “freezes” the reading and stops all further measurement updates. 

  Pressing the HOLD/MAX button again will initiate the MAX mode. The "MAX" 

annunciator appears in the upper right hand corner of the display. 

  In the MAX mode, the thermometer automatically displays and updates the maximum 

absolute values. 

  To exit the MAX mode, press the MAX button again. The Model CA863 will be in a 

normal (non MAX or HOLD) mode at power up. 

www.