16
Digital Flexprobe
®
Models 400D & 4000D
4.6 Environmental
Operating Temperature:
32° to 122°F (0° to 50°C)
Storage Temperature (without batteries):
-4° to 158°F (-20° to +70°C)
Operating Relative Humidity:
80% RH to 122°F (50°C)
Storage Relative Humidity:
90% RH up to 113°F (45°C)
The sensor can withstand a temperature of 194°F (90°C)
Indoor use / Level of pollution: 2 / Altitude: < 2000m
4.7 Mechanical
Dimensions:
3.94 x 2.36 x 0.79" (100 x 60 x 20mm)
Cable Length:
6 ft (1.83m)
Sensor Length:
400D-6: 6" (170mm)
400D-10: 10" (250mm)
4000D-14: 14" (350mm)
4000D-
2
4: 24" (610mm)
Sensor Diameter:
400D-6: Ø 1.77" (45mm)
400D-10: Ø 2.75" (70mm)
4000D-14 : Ø 3.94" (100mm)
4000D-24 : Ø 8" (190mm)
Weight:
0.29 lbs (130g) approx (MiniFlex)
Index of protection:
IP 40 per IEC 60529
IK 04 per IEC 50102
V0 (per UL 94)
The flexible coil is resistant to oils and aliphatic hydrocarbons.
4.8 Safety
Electrical safety per IEC 61010-2-032 for type B sensors
Rated 600V CAT IV
4.9
Electromagnetic Compatibility
Emissions and immunity in an industrial setting compliant with IEC
61326-1 for portable devices.