Instrument Configuration

NOTE: 

The instrument configurations can also be modified through the DataView

®

 software.

 

NOTE: 

The instrument must be configured the first time it is used. The configuration is 

saved in memory when the instrument is turned OFF. 

Press the 

 button to configure the unit. The following sub-menus appear:

•  Set the display language by pressing the yellow button corresponding to the screen  

language icons.

•  The parameter that is ready to be configured will be highlighted in yellow. To move to a 

different parameter, use the

 ▲ 

and

 ▼ 

buttons.

•  Press the Enter 

 button to select a parameter.

•  Use the 

◄ 

and

 ► 

buttons to change a value or setting.

•  When finished, return to the Configuration menu by pressing the 

 button.

PARAMETER

FUNCTION

Date / Time

Sets the date and time format

Display

Adjusts the contrast and brightness of the display; 

Defines the color of the voltage and current curves

Calculation Methods

Determines if harmonics are used or not used in calculations of reactive quantities  

(power and energy)

•  With harmonics: 

Harmonics are taken into account when calculating reactive 

parameters.

•  Without harmonics:

 Only the fundamental part is used for the calculation of  

reactive parameters

Electrical Connection

Determines the type of connection to the network

•  Single-Phase

•  Split-Phase

•  3-Phase 4-Wire

•  3-Phase 5-Wire

Sensors & Ratios

Defines the type of current probe to connect

•  MN93:  200A

•  MN193: 100A or 5A (with variable ratio)

•  SR193: 1000A

•  SL261: 10A and 100A range

•  AmpFlex

®

 Sensors: 3000A (measures up to 6500Arms)

•  MiniFlex

®

 Sensors: 1000A

•  MR193: 1000A

AC

/1200A

DC

•  5A three-phase adapter (3-channel only)

Transient Mode

Configures the voltage and current thresholds

Trend (Recording) Mode

Selects the parameters to record (up to 4 configurations)

Alarm Mode

Defines the parameters of an alarm

Erase Memory

Deletes configurations, alarm settings, snapshots and recordings

About

Displays the serial number, software and hardware version

Содержание 2136.20

Страница 1: ...QUICK START USER GUIDE 8335 PowerPad IMPORTANT WARRANTY NOTE By registering online within 30 days from the date of purchase your warranty will be extended to 3 years...

Страница 2: ...ent to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibratio...

Страница 3: ...aView User Manual NiMh Battery installed Kits will also include order dependent 4 MN93 BK current probes 4 MN193 BK current probes 4 SR193 BK current probes 4 AmpFlex 193 24 BK 24 current sensors 4 Am...

Страница 4: ...input Harmonics Mode Displays the harmonics in percent and value ratios for voltage current and power for each harmonic through the 50th Determines harmonic current produced by non linear loads Analyz...

Страница 5: ...voltage inputs 7 USB port 8 Input for external power supply and battery charging 9 Confirm Enter button 10 Navigation buttons 11 Six mode buttons see chart left Connection Terminals 3 1 2 1 Four 4 cu...

Страница 6: ...voltage and current curves Calculation Methods Determines if harmonics are used or not used in calculations of reactive quantities power and energy With harmonics Harmonics are taken into account when...

Страница 7: ...2 L3 and N Disconnecting Proceed in the reverse order to connecting always finishing by disconnecting the ground and or neutral when distributed Disconnect the leads and press the button to turn the i...

Страница 8: ...older view to launch the Dataview setup program NOTE For more information on using DataView refer to the Model 8335 user manual that is supplied on the USB stick Updating Software Firmware To provide...

Страница 9: ...certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 60...

Страница 10: ...he one year period you may return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or r...

Страница 11: ...NOTES...

Страница 12: ...10 13 99 MAN 100392 v2 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: