OPERATION 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model MN09 

•  Connect the black lead of the current probe to “common” and the red lead to the DC voltage input on

your DMM or other voltage measuring instrument. The AC current probe has an output of 100mV DC/A

AC. This means that for 150A

AC

 in a conductor around which the probe is clamped, 15V

DC

 will come out

of the probe leads to your DMM or instrument. Select the range which corresponds to the measured

current. If the current magnitude is unknown, start with the highest range and work down until the

appropriate range and resolution are reached. Clamp the probe around the conductor. Take the reading

on the meter and multiply it by 10 to obtain the measured current (e.g, 1000mV

DC

 reading = 1000 x 10 =

10,000mA

AC

 or 10A

AC

).

• For best accuracy, avoid if possible, the proximity of other conductors which may create noise.

Tips for Making Precise Measurements

•  When using a current probe with a meter, it is important to select the range that provides the best

resolution. Failure to do this may result in measurement errors. 

• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air

gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for

power measurement.

MAINTENANCE: 
Warning

 

•  For maintenance use only original factory replacement parts.
•  To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•  To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents

into the probe.

Cleaning 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 

Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 

600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.

REPAIR AND CALIBRATION 

You must contact our Service Center  for a Customer Service Authorization number (CSA#). This will 

ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# 

on the outside of the shipping container.  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

15 Faraday Drive • Dover, NH 03820 USA 
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • 

[email protected]

(Or contact your authorized distributor.) 

NOTE: All customers must obtain a CSA# before returning any instrument. 

TECHNICAL AND SALES ASSISTANCE 

If you are experiencing any technical problems, or require any assistance with the proper use or 

application of this instrument, please contact our technical hotline: 

(800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) • 

[email protected]

LIMITED WARRANTY 

The current probe is warranted to the owner for a period of two years from the date of original purchase 

against defects in manufacture. This limited warranty is given by AEMC

®

  Instruments, not by the 

distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused 

or if the defect is related to service not performed by AEMC

®

 Instruments.  

Full warranty coverage and product registration is available on our website at: 

www.aemc.com/warranty.html. 

Please print the online Warranty Coverage Information for your records. 

Отзывы: