Product model

 

The 5070-

ABE is indicated for use with the Philips™ HeartStart™ OnSite/Home, models M5066A, M5068A, 

and the FRx, model 861304. 

Please see the Philips™ HeartStart™ OnSite/Home, models M5066A, M5068A, and the FRx, model 861304 
Owner’s Manual for complete information on AED use.

 

Indications for use 

The automated external defibrillator (AED) battery supplies power to an AED as required during self 
maintenance, automated diagnoses, and defibrillation.  The 5070-ABE is indicated for use with the 

Philips™ 

HeartStart™ OnSite/Home, models M5066A, M5068A, and the FRx, model 861304.

 

Contraindications 

Automated external defibrillators should not be used when a patient is conscious or breathing normally. 

Battery specifications 

Battery Type: 9VDC, 4.2Ah, lithium manganese dioxide, primary cells. 

 

Lithium Content: approximately 5.04 g. 

 

Capacity: When new a minimum of 200 shocks or 4 hours of operating time at 25°C (77°F). Actual battery 
operating life depends on device settings, usage, and environmental factors. 

 

 

Charge Time: A new battery typically takes 8 seconds to charge the AED to maximum energy.  Batteries 
with reduced capacity will require additional time to charge the AED. 
 

Shelf Life (prior to insertion): A minimum of 5 years from date of refurbishing when stored from 20° to 30°C (68° 
to 86°F), 10% to 75% RH (non-condensing). 

 

Standby Life (after insertion/between use): 4 years when operated from 20° to 30°C (68° to 86°F), 10% to 75% 
RH (non-condensing). 

 

Operating Conditions: 0°C to 50°C (32° to 122°F), 0% to 95% RH (non-condensing). 

 

Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-4 

 

Enclosure Protection: IP55 when housed in AED. 

Transportation: UN38.3 tested: T1 

– Altitude, T2 – Thermal, T3 – Vibration, T4 – Shock, T5 – External Short 