20
8.1.3 Power Induced Failure
Application of power with poor start up characteristics can cause the
internal controller to malfunction. This is not actually a contrast problem, but
the faint blue display is a symptom of the power start up malfunction. You
need to remove power for about 5 seconds, then restart the unit. If there are
no batteries installed, just unplug the power pack. If batteries are installed,
the battery power must be interrupted using one of two methods:
1. Remove battery cover, then remove at least one battery. Replace
battery after 5 seconds, then replace cover.
2. Plug power pack into unit, but do not plug the power pack into the
wall. If you are using wall power, plug the power pack into the wall
after 5 seconds. If you are using battery power, unplug the power
pack from the Bravo MRI II unit.
The faint blue rectangle should be gone. After the power has been
removed for 5 seconds, you may press the ON key to restart the unit.
If this is a recurring problem, you should investigate your power source,
i.e. weak batteries, poor wall power, damaged power pack, etc.
8.2 Batteries
Install a fresh set of "AA" batteries; or (if possible) observe the operation
of the Bravo MRI II while switching back and forth between battery power
and and external DC supply (refer to Section 0).
8.3 Serial Port
Ensure that your cabling and communication software is configured in
accordance with the specification in Section 9.1.4.2.
VIA-RTD Users: Verify that your Software and Computer settings
conform with the specifications in Section 9.1.4.2.
Terminal Users: Verify that you are observing the protocols
specified in Sections 9.1.4.2 and 9.3