ADWA AD 310 Скачать руководство пользователя страница 9

9

1. DIN connector for conductivity probe
2.

ON/OFF

 key, to turn the instrument ON and OFF

3.

RANGE/FIXED

 key, to select measurement unit,

and to freeze current range on the LCD

4.

Up arrow/SETUP 

key,

 

to manually increase the

value of temperature or other parameters, and to
enter/exit setup mode

5.

Down arrow/CFM

 key, to manually decrease the

value of temperature or other parameters, and to
confirm values

6.

HOLD/CAL

 key, to freeze the reading on the LCD,

and to enter/exit calibration mode

7.

SHIFT

 key, to activate the key alternate function

(press and hold first the SHIFT key and then the
second desired key)

8. Liquid Crystal Display (LCD)

20

• When the reading is stable, the hourglass tag stops

blinking. Press SHIFT and CFM keys to confirm
calibration.

• The instrument displays the “Str” message and returns

to measurement mode.

Notes:

• The calibrated custom value is considered the

value of the calibration solution at the selected
reference temperature.

It is possible to set the cell constant value directly,
without following the calibration procedure. To
set the cell constant enter the setup mode and
select the “CEL” item (see “Setup” section for
details).

• The temperature reading is not used during custom

calibration.

Содержание AD 310

Страница 1: ...EC TDS and Temperature w w w a d w a i n s t r u m e n t s c o m ADWA INSTRUMENTSKft Als kik t sor 11 6726 Szeged Hungary Tel 3662317878 Fax 3662550610 e mail sales adwainstruments com www adwainstru...

Страница 2: ...mmences from the original date of sale Warranty is only valid when the product is used under normal conditions and in accordance with the instruction manual The warranty is void if the instrument is r...

Страница 3: ...ent 25 Probe Maintenance 26 Probes and Solutions 27 TABLE OF CONTENTS 26 PROBE MAINTENANCE After measurements rinse the probe with clean water If a more thorough cleaning is required remove the probe...

Страница 4: ...ser selectable The instruments also feature a measurement stability indicator 25 BATTERY REPLACEMENT When batteries become weak the battery symbol lights up It is recommended to replace the batteries...

Страница 5: ...ation solution 20 ml sachet each 1413 S cm for AD310 1382 ppm for AD410 Batteries 4 x 1 5V AA alkaline Instruction manual 24 To freeze the reading on the LCD press the HOLD key from measurement mode T...

Страница 6: ...ne custom value TDS Calibration Offset 0 0 ppm AD410 Slope 1 point with 4 memorized standards 800 1382 ppm 6 44 40 0 ppt or one custom value 23 Note Press SHIFT and then SETUP key before confirmation...

Страница 7: ...vironment 0 to 50 C RH max 95 Dimensions 70 x 145 x 26 mm Weight Approx 200 g 22 SETUP Setup mode allows to view and modify the instrument parameters To enter setup mode press SHIFT and then SETUP key...

Страница 8: ...laythefactorydefaulttemperature Adjust the temperature reading on the LCD using the arrow keys The maximum allowed adjustment is 1 0 C around current reading Press SHIFT and CFM keys to confirm The me...

Страница 9: ...tivate the key alternate function press and hold first the SHIFT key and then the second desired key 8 Liquid Crystal Display LCD 20 When the reading is stable the hourglass tag stops blinking Press S...

Страница 10: ...or a few seconds then the LCD displays the not calibrated temperature compensated EC or TDS reading Pressing SHIFT and RANGE keys the LCD will show the temperature compensated EC or TDS reading factor...

Страница 11: ...theair Notes If temperature is out of range the buF message and the C tag blink on the LCD If the reading is too far from the expected value buF message blinks For best results calibrate using a stand...

Страница 12: ...rature reading in the selected range Notes If the meter displays only dashes the reading is out of range Ifthestabilityindicator hourglasssymbol blinks the reading is not stable 17 If necessary use th...

Страница 13: ...t samples for accurate reading it is recommended to rinse the probe thoroughly with deionized water before immersing it into the sample 16 CALIBRATION STANDARD CALIBRATION EC AD310 or TDS AD410 calibr...

Страница 14: ...ompensation setting can be accessed by entering the setup mode and selecting the tcE item see Setup section for details 15 If the temperature compensation is active measurements are compensated using...

Отзывы: