• The hourglass tag will blink on the LCD until the read-
ing is stable.
• When the reading is stable and close to
the selected buffer, the CFM tag ap-
pears. Press the CFM key to confirm.
• The calibrated value is then displayed on the primary
LCD and the secondary LCD will display the second
expected buffer value, together with CAL and “Cal
Point 2” tags.
• After the first calibration point is confirmed, immerse
the pH electrode and the temperature probe approxi-
mately 4 cm into the second buffer solution and stir
gently. The temperature probe should be close to the
pH electrode.
• If necessary, use the arrow keys to select a different
buffer value.
• The hourglass tag will blink on the LCD until the read-
ing is stable.
24
45
Notes:
• The custom buffers can be set only with 0.001
resolution. If 0.01 pH resolution is selected dur-
ing calibration, the displayed custom buffer value
will be a rounded one.
• While in setup changing mode for items calibra-
tion alarm time-out, custom buffer and alarm lim-
its, press CLEAR to set the selected item to the
default value.
• Refer to the table below to select the correct ion
charge for each ion type (AD1200 only):
ION CHARGE
ION TYPE
-2 (divalent anion)
S, CO
3
-1 (monovalent anion)
F, Cl, Br, I, CN, SCN, ClO
4
, NO
3
+1 (monovalent cation)
H, Na, K, Ag, NH
4
+2 (divalent cation)
Mg, Ca, Ba, Cd, Cu, Pb
undf
undefined ion