Guidance
and
manufacture
’
s
declaration
–
electromagnetic
immunity
The
Arm
Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of
Arm
Blood Pressure Monitor should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Complia
nce
level
Electromagnetic
environment
-
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not applicable
Portable and mobile RF communications
equipment should be used no closer to any
part of the Wrist Blood Pressure Monitor,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power
ratingof the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m