3.2
Internal Data Flow
3-2
3.2.
Internal Data Flow
In the R3755x/R3760, basic calculations (port calculation, calibration and parameter conversion) are per-
formed based on the data input to R-ch and A-ch. After which, LOGMAG calculations and phase calculations
are also performed (refer to
). For other formats, calculations must be performed at application level
using unformatted data (Real/Imag).
Port
Raw data
R input
A input
calculation
R, A, A/R
Port1 input S11.S21
Port2 input
Figure 3-2 Internal Data Flow
data
LOGMAG
data
PHASE
data
Parameter
conversion
Z, Y, l/S
Data
Calibration
data
Error
correction
·
Normalization
·
Full calibration
Port extension
Electrical
length
correction