R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
2.3.9 Harmonic Distortion Measurements
2-108
2.3.9.2 Using the Peak List
This section describes a method on how to measure harmonic distortion using the peak list.
Measurement conditions: The target of the measurement below is a signal that has an output frequency
of 100 MHz and a level of -10 dBm.
Use appropriate parameter values when making the measurements in the
example shown below.
Setup
1.
Connect the unit under test as shown in Figure 2-121.
Figure 2-121 Setup for Measuring the Harmonic Distortion
Power on
2.
Turn the analyzer and the signal generator power on.
Setting the signal generator
This prepares the signal generator for output.
3.
Set the frequency to 100 MHz; the level to -10 dBm; and output to ON.
Initialization
This resets the current settings to the factory defaults or user-defined presets.
4.
Press SHIFT and CONFIG(PRESET).
This sets the analyzer to its presets values.
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
5.
Press FREQ, Start, 5, 0 and MHz.
The start frequency is set to 50 MHz.
6.
Press Stop, 3, 5, 0 and MHz.
The stop frequency is set to 350 MHz.