R3132 Series Spectrum Analyzer Operation Manual
5.3.3 Output Level Switching Accuracy
5-97
Table 5-28 TG Output Level Switching Accuracy Test Setting (1)
11. Repeat 2 through 9 for each center frequency listed in Table 5-29.
Center Frequency
Output Level
(dBm)
Reference Level
(dBm)
30 MHz
-10.0
-5.0
0.0
5.0
-14.9
-9.9
-15.0
-10.0
-19.9
-14.9
-20.0
-15.0
-30.0
-25.0
-40.0
-35.0
-50.0
-45.0
Table 5-29 TG Output Level Switching Accuracy Test Setting (2) (1 of 5)
Center Frequency
Output Level
(dBm)
Reference Level
(dBm)
100 kHz
-10.0
-5.0
0.0
5.0
-14.9
-9.9
-15.0
-10.0
-19.9
-14.9
-20.0
-15.0
-30.0
-25.0
-40.0
-35.0
-50.0
-45.0