EMC

This function measures electromagnetic interference generated
by various electronic equipment. This function incorporates
the 9 kHz and 120 kHz RBW and QP detector conforming to
the CISPR Pub.16-1 standard. In addition, using the AM/FM
demodulation signal fed from the PHONE jack on the rear
panel, you can identify broadcasting radio waves which act as
external noise. Prior to measurement of noise emission on the
approved site, this function is very useful for preparatory
evaluation and solution.

Various antenna correction factors provided by Advantest are
built-in the R3131A. Simply by selecting the Model name of
the antenna, the level indication of the R3131A is calibrated to
an absolute value, allowing you to read the value directly in

unit of dBµ/m. When you use an antenna from other
manufacturers, you can reflect its antenna correction factor in
the level indication of the R3131A by performing steps

and

below.

Diverse measurement functions

The MEAS key incorporates the XdB Down measurement
function which is useful for noise measurement, AM
modulation measurement, 2-signal 3rd-order distortion
measurement, and filter cut-off frequency measurement. In
noise measurement, bandwidth conversion can easily be made
and the PBW calibration function for improvement of
measurement accuracy is effective. The PBW calibration
function is a new calibration function which performs
correction, in power measurement, based on conversion of the
R3131A resolution bandwidth filter to an ideal filter, thereby
allowing measurement with higher accuracy.

7

R3131A-4E Jan. ´02

Antenna and Level Correction Functions

(EDIT of Corr.table is not performed by the R3131A unit.)

SHIFT

RECALL

SAVE

Save

Save Item

Protect

Delete

1.

Set Antenna Correction to ON and SAVE the file.

2.

OPEN the file from the floppy disk using Excel 

    on the PC.

3.

Enter the frequency and correction level in 

    the 

ANT CORR

 area and then overwrite it 

    on the floppy disk.

4.

Load the floppy disk in the R3131A and then 

    RECALL the file. The Correction table is created.

Setup ON/OFF

Trace ON/OFF

Antenna

Correction 

ON

/OFF

MAX. 50 Points

1

2

Set Correction to ON. The corrected data is reflected 
on the screen data.

SHIFT

1

EMC

Field

Correction 

ON

/OFF

PERSONAL SPECTRUM ANALYZER

Содержание R3131A

Страница 1: ...R3131A R3131A Spectrum Analyzer A personal spectrum analyzer for use in diverse applications...

Страница 2: ...l Layout Data Entry Keys The data entry keys arranged together with the FREQ SPAN and LEVEL basic functions improves operability Marker Keys Various marker functions like delta marker and peak search...

Страница 3: ...interfaces GPIB RS232C Centronics and FD drive Large character display allows results to be seen Substantial EMC measurement function Improved system operation speed Operation key arrangement for eas...

Страница 4: ...existed immediately before execution of AUTO TUNE allowing observation under the same measurement conditions COUNTER Performs frequency measurement with the built in frequency counter simply by movin...

Страница 5: ...display it as the channel power simply by setting the measurement window to the specified occupied bandwidth Total Power Obtains the total power from the spectrum displayed on the screen This function...

Страница 6: ...od in which either a signal level or frequency is evaluated for wave shape in accordance with a given window range The other is a limit line method in which electromagnetic compatibility EMC and filte...

Страница 7: ...n which is useful for noise measurement AM modulation measurement 2 signal 3rd order distortion measurement and filter cut off frequency measurement In noise measurement bandwidth conversion can easil...

Страница 8: ...e conditions Use of Data Save Recall of the R3131A 8 R3131A 4E Jan 02 Series of center frequency sweep marker and data load operations Former products from Advantest Former products from Advantest For...

Страница 9: ...system highly accurate measurement is possible Because the output level can be set in a wide range from 0 to 59 9 dBm in 0 1 dB steps it can be used to measure filter pass characteristic cable loss a...

Страница 10: ...0 dBm 0 3 dB Frequency response 0 5 dB 100 kHz to 3 GHz ATT 10 dB 1 dB 100 kHz to 2 GHz 2 dB 9 kHz to 3 GHz after calibration at 30 MHz reference Scale display accuracy LOG 0 5 dB 0 to 20 dB after aut...

Страница 11: ...eep R3131A Spectrum Analyzer OPT 74 Tracking generator option 10MHz REFERENCE INPUT PRINTER PCL ESC based PHONE PROBE POWER GPIB Remote Data Out Applicable printer control code ESC P ESC P Raster HP P...

Страница 12: ...k Singapore 119967 Tel 65 274 3100 Fax 65 274 4055 Tektronix Inc North America P O Box 500 Howard Vollum Industrial Park Beaverton Oregon 97077 0001 U S A Tel 1 800 426 2200 Fax 1 503 627 4090 Rohde S...