47
Appendix D
5.
At this step, the Wizard will output –10V on the channel you
selected. Please check the value on the meter is in the range (
I
). If
the value does not fit, please perform the adjustment.
There are 3 ways to perform the adjustment.
1. Move the slider (II).
2. Input the adjust value (III). Note that only if the [Set] button is
pressed, the value will take effect.
3. Press [Default] button (IV). This will load the factory default
value.
The value on the meter will change after you modify the Adjust
Value. Please check the value on the meter is in the range again. If
value fits the condition, press the [Next] button to continue.
Figure D.4: The Adjustment Process of Offset Calibration