Chapter 1
Installation and setup
19
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
1.5 Connecting to a network (optional)
To connect the OsmoTECH XT to a laboratory
information system (LIS), a local area network (LAN), or
to enable the web server, plug an Ethernet cable
connected to your network into the Ethernet port on the
back of the OsmoTECH XT.
You must also configure the communication settings in
the OsmoTECH XT software. For more information, see
Setting up networking communications
.
M
J
K
L
N
J.
USB 2.0 Type A ports (2)
K.
USB 2.0 Type B port
L.
Ethernet port
M.
Power switch
N.
Plug for power cord
Powering up the instrument
1.
Connect the power cord to the back of the instrument
and insert the plug into the power outlet.
2.
Turn on power to the instrument using the rocker
switch on the back. Depress the side marked with the
|
symbol.
When the instrument powers up, the splash screen .
3.
After a few moments, the OsmoTECH XT Home
screen opens.
From the Home screen, you can initiate sample
testing and view the results. Initially, the instrument
status shown in the green status banner is
Ready
.
4.
If one of the following messages appears after
starting up, follow the associated instructions:
•
Preventative Maintenance Required
- An Advanced
Instruments trained service engineer must perform
any scheduled maintenance and record it on the
DIAGNOSTICS > MAINTENANCE screen (see
Logging preventative maintenance
).
•
Installation date not se
t.
Refer to the User Guide
- Follow the instructions in the next section,
Setting
the installation date
.
Setting the installation date (Administrators only)
If the installation date of the instrument has not been
entered into the system, a
Installation date not se
t.
Refer
to the User Guide
message appears the first time an
instrument powers up. If the installation date is not set,
this message reappears every hour.
Complete the following steps to enter the installation
date to disable this message and enable the preventative
maintenance reminders based on this date.
1.
Press
> DIAGNOSTICS > MAINTENANCE
and log
in.
2.
Press the
INSTALLATION t
ab.